Trouble Shooting

WARNING TROUBLE SHOOTING SHOULD ONLY BE CARRIED OUT BY

ALLEGRO CONCEPTS AUTHORISED PERSONNEL FAMILIAR WITH

BOTH USE, ELECTRICAL AND MECHANICAL FEATURES OF PATIENT

LIFTING HOISTS.

The following is a guideline only to general trouble shooting. Before carrying out any trouble shooting 

ensure that.

• 

Red emergency stop button is released.

• 

Maintenance procedures checked & followed.

• 

All cable connections are secure.

• 

Battery is fully charged.

POSSIBLE FAULT

ACTION

Warning buzzer sounds

within control box.

Flat battery pack.

Faulty battery pack.

Recharge battery.

Replace battery.

Battery charger power indicator

LED does not light up.

No mains connection.

Internal fuse blown.

Battery chargers is defective.

Plug into mains.

Replace fuse.

Repair/replace charger.

Actuator will not run and no 

relay

click is heard from control box.

Control box is defective.

Handset defective.

Replace control box.

Actuator will not run via handset

but control box relays click.

Actuator plugs not fully pushed

in to control box.

Actuator defective.

Control box defective.

Push fi rmly into correct port.

Replace actuator or control.

Actuator cannot lift full load or

runs slow.

Batteries not charged/defective.

Load too high for lifter.

Actuator motor damaged.

Charge/replace batteries.

Replace actuator.

No response from control box,

no clicking of relays is heard.

Emergency stop button 

engaged.

Hoist battery completely fl at.

Blown fuse in battery pack.

Faulty control box.

Release.

Charge battery.

Replace/repair control box.

Actuator works in one

direction only.

Handset defective.

Control box defective.

Replace handset.

Repair/replace control box.

Содержание SALSA 200

Страница 1: ...SALSA Mobile Standing Transfer Hoist Instruction Safety Manual Please read this manual before operation for maintenance instructions and safe usage The enabling power of applied knowledge...

Страница 2: ...u W www unicarehealth com au Contents Overview 2 Important Saftey Considerations 3 Assembly Instructions 5 Hoist Operation 6 Actuator 6 Charging 7 Charging Indicator Lights 7 Charging Do s and Don ts...

Страница 3: ...d to both home or institutional use The SALSA is an electric operated hoist with ergonomic features designed with the patient and carer in mind An ALLEGRO standing sling speci cally designed for the S...

Страница 4: ...ng task should be assessed planned by a trained user Safety Considerations Ensure all that all users read this guide before operating the hoist Do not operate the hoist without training in the safe op...

Страница 5: ...4 T 08 9351 7500 E info unicarehealth com au W www unicarehealth com au Lift Arms Actuator Knee Pad Leg Battery Pack Push Handle...

Страница 6: ...nut 3 Repeat process 2 for second leg 4 Attach leg adjuster tie rod ends to each leg and tighten each nylock nut 5 Attach lift actuator to lifting arm assembly by aligning actuator clevis with actuato...

Страница 7: ...arrows and symbolised legs indicating the direction of travel This function allows an unrestricted range of leg adjustment The handset is also tted with a battery indicator light This will ash when th...

Страница 8: ...from control box by pulling the handle on the pack towards you Use both hands to support the battery pack when removing or replacing it Clip the battery pack onto the charging unit ensuring it is seat...

Страница 9: ...ty The batteries are not the nickel cadmium type and must not be periodically fully discharged Battery life is greatly reduced by deep or complete discharging of the batteries Longer lifetime is obtai...

Страница 10: ...ance is suf cient to manoeuvre legs under the bed Lower the SALSA lift arms using the handset buttons until sling keyhole buckles can reach and easily engage the studs on the lift arms to prevent poss...

Страница 11: ...gue lubricate and or adjust as required NB leg pivot bushes incorporate a Te on coated pivot bush which requires minimal lubrication Check tightness of bolt in leg bushes Check linkages are secure in...

Страница 12: ...up No mains connection Internal fuse blown Battery chargers is defective Plug into mains Replace fuse Repair replace charger Actuator will not run and no relay click is heard from control box Control...

Страница 13: ...KG BASE LENGTH 1050mm EXTERNAL BASE WIDTH 655mm INTERNAL BASE WIDTH CLOSED 585mm INTERNAL BASE WIDTH OPEN 1035mm BASE WIDTH ADJ ELECTRIC UNDER BED CLEARANCE 125mm MIN HEIGHT OF LIFT ARM 825mm MAX HEIG...

Страница 14: ...the improper or unsafe use of the product No warranty claim shall apply where the product or any part thereof has been modi ed varied altered or damaged either accidentally or through improper or negl...

Страница 15: ...7500 E info unicarehealth com au W www unicarehealth com au Log Book Service Type Condition Report Date Inspection by Signature Action Taken Service Type Condition Report Date Inspection by Signature...

Страница 16: ...PO Box 82 Bentley WA 6982 T 08 9351 7500 F 08 9350 5299 info unicarehealth com au www unicarehealth com au Authorised Distributor...

Отзывы: