Trouble Shooting
WARNING TROUBLE SHOOTING SHOULD ONLY BE CARRIED OUT BY ALLEGRO CONCEPTS
AUTHORISED PERSONNEL FAMILIAR WITH BOTH USE, ELECTRICAL AND MECHANICAL
FEATURES OF PATIENT LIFTING HOISTS.
The following is a guideline only to general trouble shooting.
Before carrying out any trouble shooting ensure that.
•
Red emergency stop button is released.
•
Maintenance procedures checked & followed.
•
All cable connections are secure.
•
Battery is fully charged.
POSSIBLE FAULT
ACTION
Warning buzzer sounds
within control box.
Flat battery pack.
Faulty battery pack.
Recharge battery.
Replace battery.
Battery charger power indicator
LED does not light up.
No mains connection.
Internal fuse blown.
Battery chargers is defective.
Plug into mains.
Replace fuse.
Repair/replace charger.
Actuator will not run and no
relay
click is heard from control box.
Control box is defective.
Handset defective.
Replace control box.
Actuator will not run via handset
but control box relays click.
Actuator plugs not fully pushed
in to control box.
Actuator defective.
Control box defective.
Push fi rmly into correct port.
Replace actuator or control.
Actuator cannot lift full load or
runs slow.
Batteries not charged/defective.
Load too high for lifter.
Actuator motor damaged.
Charge/replace batteries.
Replace actuator.
No response from control box,
no clicking of relays is heard.
Emergency stop button
engaged.
Hoist battery completely fl at.
Blown fuse in battery pack.
Faulty control box.
Release.
Charge battery.
Replace/repair control box.
Actuator works in one
direction only.
Handset defective.
Control box defective.
Replace handset.
Repair/replace control box.