5
KRATOS 320 Mobile Patient Lifting Hoist Instruction & Safety Manual
T
08 9351 7500
E
W
www.unicarehealth.com.au
Assembly Instructions
The ALLEGRO range of patient lifting and transfer hoists are fully assembled, load tested and certifi ed
at the factory before being partially dismantled for packing and shipping.
Use the instructions below as a guide to assemble your KRATOS 320 hoist.
1. Remove lifter components from protective packing and lay out on a smooth surface to prevent
damage to powder coating on lifter sub assemblies.
2. Lift the mast assembly and slide into mast holder receiving socket on base chassis. Ensure the mast
assembly is completely engaged in the mast socket. Insert the locating bolt through the chassis and
base and tighten the locking levers to prevent movement between mast and base chassis.
3. Attach lift actuator to jib, by aligning actuator clevis with actuator mount bracket. Insert bolt and
tighten nylock nut.
4. Attach push handle assembly with bolts supplied with handle assembly.
5. Attach battery by clipping battery pack on top of control box.
6. Plug the lift actuator cable into port 1 positioned on the underside of the control box, ensuring it is
driven fi rmly home in the socket.
7. Your KRATOS 320 hoist uses an electric leg adjustment; plug the leg actuator lead into port 2
positioned on the underside of the control box.
8. Plug the handset lead into the large socket on underside of control box. (See next diagram) ensure
alignment of tongue on handset plug with groove in control box handset connection port. When
aligned press home fi rmly. A smear of Vaseline on the o ring may assist if plug is tight.
9. Check operation of lifter via handset, ensuring that the RED EMERGENCY STOP BUTTON ON
THE CONTROL BOX IS RELEASED. Turn the knob quarter turn clockwise to release.