background image

18

T

 08 9351 7500              

E

 [email protected]              

W

 www.unicarehealth.com.au

Use of the Hand Control
Session Time

This function shows the running time (in minutes) of the session (motor functioning).

•  It directly accessible by the key,          the display shows

This counter is reset each time the unit is switched ON.

TIME

  30 STOP

02H25

15   90

Содержание Kinetic Spectra

Страница 1: ...Kinetic Spectra Instruction Safety Manual Please read this manual before operation for maintenance instructions and safe usage The enabling power of applied knowledge...

Страница 2: ...Technical speci cations 23 Symbols used 24 Warranty 24 Maintenance sheet 25 Safety Recommendations The physician de nes the protocol and ensures that it is correctly implemented adjustments session ti...

Страница 3: ...in ammation and the loss of range of motion Prevents joint stiffness Speeds the reco very of post operative range of motion Maintains the quality of the joint surface Reduces pain and edema Promotes...

Страница 4: ...t 4 Hand control 5 Thigh support setting lock 6 Lower limb support setting lock 7 Foot support positioning setting lock 8 Transport handle 9 ON OFF switch and fuses 10 Liquid crystal display 2 lines o...

Страница 5: ...se function on the control handle Hand control must be accessible to patient at all times See page 7 KINETEC Spectra complies with Directive 93 42 CEE EXPLOSION HAZARD KINETEC Spectra is not designed...

Страница 6: ...rical socket is in good condition and is suitable for the splint power supply cord The latter complies with current standards and has a grounding socket The plug may be connected to any standard socke...

Страница 7: ...if the hand Please wait control is not blocked Kinetic See page 7 30 STOP 15 90 Press the 2 keys in Language The display indicates the same time English the language selected To change the Language T...

Страница 8: ...press the and keys The display reads UNLOCK We recommend that you lock the hand control when you give it to the patient Comment The hand control is unlocked when you switch the unit ON OFF START STOP...

Страница 9: ...ch OFF 30 STOP 15 90 this mode by pressing on the key Start the session with WARMUP The value change the original 30 RUN 45 90 at the speed of the Parameters of the movement movement default setting P...

Страница 10: ...NETIC The new value blinks if necessary 40 Flex 15 110 To validate the new KINETIC While the value value press another SPEED 2 blinks press key or wait more than FORCE the 4 seconds No timer to change...

Страница 11: ...g the same time and Spectra V01 0 3 seconds switch the unit ON Then Program 2 The program number EMPTY blinks To change the program Program 10 The new program if necessary 25 kinetic 110 number blinks...

Страница 12: ...SPEED 2 press the key or wait more FORCE No timer to change if Pause low 0 necessary To record the program 10 PROGRAM 10 Save clear Then PROGRAM 10 The program 10 has Save been recorded and PROGRAM 1...

Страница 13: ...ehealth com au Use of the Hand Control Comments When a program has been deleted the display shows Program table Program Treatment Flexion Extension Speed Force Flexion Extension Timer Number Mode Limi...

Страница 14: ...e 7 To access the program PROGRAM 1 The program number mode 70 WARMUP 10 blinks To change the program PROGRAM 3 The new program if necessary 25 WARMUP 90 number blinks To exit and validate KINETIC The...

Страница 15: ...gram PROGRAM 1 The program number mode 70 WARMUP 10 blinks To change the program PROGRAM 3 The new program if necessary 25 WARMUP 90 number blinks To read the speed PROGRAM 3 Displays the speed value...

Страница 16: ...the movement value if necessary See page 9 Start the movement WARM UP The value change 15 RUN 15 110 at the speed of the movement Warm Up Rules The Kinetec Spectra starts at 70 of the full ROM increa...

Страница 17: ...inetic If the Warm Up is 30 STOP 15 90 selected switch OFF WARM UP this mode by pressing 30 STOP 15 90 on the key To select the MODUL use or The display includes MODULATION Mode 30 STOP 15 90 the keys...

Страница 18: ...5 90 To select the BYPASS BYPASS use or The display indicates mode 30 RUN 15 90 the keys used to run the machine This message is displayed 3 seconds To select the NEW BYPASS use or The unit is moving...

Страница 19: ...lth com au Use of the Hand Control Session Time This function shows the running time in minutes of the session motor functioning It directly accessible by the key the display shows This counter is res...

Страница 20: ...th a llabell tto rreccorrd tthe pattiientt ss name CLEANING Sterilization of the pads if necessary Sterilized at 134 C during 18 minutes Disinfecting of the pads Washing at 30 C with use of a disinfec...

Страница 21: ...knobs 5 To install the patient on the KINETEC Spectra machine Push the foot plate 3 up to the patient and tighten the knobs 6 Adjust the plantar exion 40 or the dorsal exion 30 of the foot with the k...

Страница 22: ...quest from your KINETEC distributor If after connecting the power supply cable to the power supply and switching on KINETEC Spectra The display does not indicate any information Check that the electri...

Страница 23: ...Technical speci cations Product Weight 11Kg 24 pounds plint dimensions 94cm 37 inches x 33cm 13 inches x 33cm 13 inches Angular limits 10 to 120 Speed from 30 to 160 per minute for average femur leng...

Страница 24: ...or repair in the plant of the component or components found to be defective KINETEC guarantees its joint passive mobilization systems for 1 year against all defects of manufacture from the date of pur...

Страница 25: ...m au W www unicarehealth com au Product information Maintenance sheet Serial Number Purchasing Date Date Operation Done Technician Running Time see page 22 Exchange Parts Date Operation Done Technicia...

Страница 26: ...25 Kinetic Spectra Instruction Safety Manual T 08 9351 7500 E info unicarehealth com au W www unicarehealth com au...

Страница 27: ...26 T 08 9351 7500 E info unicarehealth com au W www unicarehealth com au...

Страница 28: ...PO Box 82 Bentley WA 6982 T 08 9351 7500 F 08 9350 5299 info unicarehealth com au www unicarehealth com au Authorised Distributor...

Отзывы: