15
Kinetic Spectra, Instruction & Safety Manual
T
08 9351 7500
E
W
www.unicarehealth.com.au
Use of the Hand Control
How to use the WARM UP key
Begining
Keys to press
Display
Remarks
To stop the unit
KINETIC
Check of the hand
40 STOP 15 100
control is not blocked
(See page 7)
To select the warmup
WARM UP
The display shows the
mode
40 STOP 15 90
default settings
(see page 8)
To change the
movement value if
necessary
(See page 9)
Start the movement
WARM UP
The value change
15 RUN 15 110
at the speed of the
movement
Warm Up Rules:
The Kinetec Spectra starts at 70% of the full ROM, increasing 5% of the range
every other cycle until the pre-set ROM is reached.
Comments:
The calculation mode used enables reaching the pre-set ROM in approximately
seven full cycles.
Example:
For a Warm Up treatment with a pre-set ROM from 0° to 100°.
The fi rst cycle starts from 15° to 85° to 15° and increase 5% each cycle.
Cycle 1 (70%)
Cycle 2 (75%)
Cycle 3 (80%)
Cycle 4 (85%)
Cycle 5 (90%)
Cycle 6 (95%)
Cycle 7 (100%)
Pre-Set ROM
50°
100°
0°