VLM1/VLM1B User Manual
18
USB Operation
The
VLM1/VLM1B
can also be controlled via a serial command from a computer’s USB port
(configured as a virtual COM port). A USB cable is included with the controller. Once the device is
connected to a computer (or USB Hub*), in most instances, the driver for the USB interface will be
installed. If this connection does not occur, a USB driver program can be downloaded from:
http://www.ftdichip.com/Drivers/CDM/CDM%20v2.12.04%20WHQL%20Certified.exe
This USB driver is also included on the Flash Drive which is included with your
VLM1/VLM1B
controller. In fact, in addition to the link above, the USB driver can be found on the same Flash Drive
that this PDF User Manual is located – This user manual is also available for download at
www.uniblitz.com
.
(*Each VLM1/VLM1 unit connected to a USB Hub will require a minimum of 0.5A each for
operation.)
Once communication is established, the shutter can be commanded to the open state by sending the
ASCII ”@” symbol. The shutter will close by sending a capital “A” symbol. See USB Commands in the
“External Input Characteristics”.
Testing the communication can be accomplished with a terminal emulation program such as Termite
3.1, found at
http://www.compuphase.com/software_termite.htm
. (Termite 3.1 is copyrighted software
that is free for personal and commercial use. You may use it and distribute it without limitations. You
may however not remove or conceal the copyright. There are no guarantees or warranties whatsoever;
use it at your own risk.)
Maintenance
Proper care and maintenance of the
VLM1/VLM1B
should be taken as with any sensitive electronic
instrument.
There are no user-serviceable parts on the
VLM1/VLM1B
. The +5VDC input on the PWR/COM
connector and the Shutter Fuse on SHUTTER connector Pin#1 are resettable.
It is highly recommended that if a problem is suspected with the unit that it be returned to the factory for
checkout, proper adjustments, and calibration. Failure to do this may damage the unit’s circuitry and/or
functionality and will void the factory warranty.
Inspection
Inspect the
VLM1/VLM1B
for damage, wear, and missing parts. If the unit appears to show signs of
damage, it should be checked thoroughly to verify correct operation and performance. If damage is
suspected, please return the unit to the factory for repair – there are no user-serviceable parts on the
board.
If there is a significant build-up of dust or other material, follow the cleaning procedure below.
Cleaning Procedure
1.
Do not use any tools to brush or scrape away accumulated materials.
2.
Do not use any liquids to clean the board.
3.
Simply use low-pressure air to blow away any dust or use a lint free cloth.
4.
There is no other cleaning required.
Содержание N-CAS VINCENT ASSOCIATES VLM1
Страница 13: ...VLM1 VLM1B User Manual 13 Figure 5 VLM1 to FS Series Shutter Interconnection Diagram ...
Страница 16: ...VLM1 VLM1B User Manual 16 Figure 6 VLM1 Dimensional Layout ...
Страница 17: ...VLM1 VLM1B User Manual 17 Figure 7 VLM1B Dimensional Layout Fuse Jumper Locations ...
Страница 22: ...VLM1 VLM1B User Manual 22 Figure 8 VLM1 VLM1B Schematic Diagram ...
Страница 23: ...VLM1 VLM1B User Manual 23 Notes ...