Start Up
After unpacking your unit inspect for any defects. Upon inspection if a visible defect is found, or a part
(or parts) are missing, notify Vincent Associates immediately.
Prior to the connection of input/output signals to the
VLM1/VLM1B
:
1.
Be sure that the power supply is disconnected.
2.
Set N.C./N.O. switch to the N.C. position (that is in the out position/LED built into the switch
actuator should be off when the supply is connected).
3.
Connect the RED wire of the FS Shutter’s actuator coil to Terminal #1 of the SHUTTER
connector and the BROWN wire of the FS Shutter’s actuator coil to Terminal #2 of the Shutter
connector. See Figure #4 for method of inserting (connecting) the actuator coil’s wires to the
VLM1/VLM1B’s
SHUTTER spring terminal block connector. If the shutter is equipped with an
electronic sync, the remaining three wires can be connected as shown in Figure #5. This figure
also shows an overview of all of the
VLM1/VLM1B
’s connections for proper operation of an FS
Shutter.
4.
When using the +5V/1A power supply connect the power supply to the
VLM1/VLM1B
with the
USB A 2.0 to USB Micro B Cable (both are included with the
VLM1).
When Using a laptop or a
computer, be sure the USB port you are using can source 0.5A minimum and then using the same cable,
connect to the
VLM1
or
VLM1B
.
5.
Plug in the power supply (If being used to power the unit.)
6.
Connect an input exposure signal, +5V TTL to the input BNC, if desired – default – Active Low.
7.
A push of the N.C./N.O. switch will cause the shutter to open. The switch push-button’s GREEN
LED will also illuminate indicating that a signal has been provided to the input switching
MOSFET amplifier. See
VLM1/VLM1B
schematic diagram in Figure #8.
Extending the Shutter Interconnection
For those applications that require an extended interconnect length between the FS shutter and the
VLM1/VLM1B
, an optional cable can be purchased, the 510FS. This cable with provide up to a 10ft
(3m) cable length between the FS Shutter and the and the
VLM1
or
VLM1B
.
The shutter must be modified to include the -5M option.
The -5M option consists of a 6-inch harness
terminated to a 5-Pin male connector added to your FS shutter. This connector will mate with the 5-Pin
Female connector on the 510FS cable. The opposite end of the 510FS contains a pigtail end that will
interconnect with the SHUTTER connector on the
VLM1/VLM1B
. The pigtail wire color layout is the
same as the colors indicated in Figure #5. It is recommended that this -5M upgrade is included with the
purchase of your FS shutter.
Be sure to observe Electro-Static Discharge (ESD) anti-static unpacking
procedures when removing the VLM1/VLM1B
from the static shielding
bag. Improper handling can result in destruction of the CMOS integrated
circuits.
CAUTION
Содержание N-CAS VINCENT ASSOCIATES VLM1
Страница 13: ...VLM1 VLM1B User Manual 13 Figure 5 VLM1 to FS Series Shutter Interconnection Diagram ...
Страница 16: ...VLM1 VLM1B User Manual 16 Figure 6 VLM1 Dimensional Layout ...
Страница 17: ...VLM1 VLM1B User Manual 17 Figure 7 VLM1B Dimensional Layout Fuse Jumper Locations ...
Страница 22: ...VLM1 VLM1B User Manual 22 Figure 8 VLM1 VLM1B Schematic Diagram ...
Страница 23: ...VLM1 VLM1B User Manual 23 Notes ...