background image

 

VLM1/VLM1B User Manual 

 

 14

Operating Basics 

Initial Operation and Testing 

The 

VLM1/VLM1B

 provides the circuitry necessary to operate an FS Series shutter from a TTL signal.   

Connect your FS Shutter to the SHUTTER output connector as described in START UP and as shown in 
Figure #5.  If your shutter is equipped with the Electronic Synchronization system, connect all five 
wires, otherwise, the two wires are all that are required to be connected. 
When the power supply cable is connected to the 

VLM1/VLM1B’s 

PWR/COM port and the power 

supply is connected to an AC source (via the AC module provided or the USB COM port of a 
computer), the Green POWER LED will illuminate.  Pushing the N.C./N.O. switch, the shutter will open 
and the GREEN LED of the N.C./N.O. switch will illuminate indicating the output MOSFET amplifier 
has been activated. 
This device also supports an FS device equipped with an Electronic Synchronization circuit. Therefore, 
when the shutter opens and if so equipped (and connected) the Yellow LED will illuminate, indicating 
the Sync system is active. Additionally, the SYNC OUTPUT BNC will change from 0V to +5V TTL 
Active High and remain high as long as the Shutter is active (Active High – default setting).  If you wish 
this output to be Active Low, reposition the internal Jumper as shown is Figure #7 on the main board. 
When the 

VLM1/VLM1B

 receives an Active Low TTL control signal (0V) on the PULSE Input BNC, 

the shutter will open and when the pulse returns to +5V the shutter will close.  The TTL control pulse 
controls a MOSFET switch that in turn controls the FS Series Shutter from the +5VDC supply. 
If the requirement is to operate the shutter from a switch contact closure, follow this procedure (this is 
when the active setting of the PULSE Input BNC is active low – the default setting): 

1.

 

Connect a wire from pin #4 of the 

SHUTTER connector

, this is the Ground connection. Or 

short the outer shell of the BNC to the center contact of the BNC.  (This is the same action as 
indicated when using the 710R or 710R/F as described below.) 

2.

 

Switch the 0V (Ground) to the BNC center contact. The shutter will open. 

The shutter will open and remain open for as long as the switch contact is depressed.  When 0V is 
removed from the input, the shutter will close.  This is the same action that will occur when the Front 
Panel pushbutton is depressed. 
After connecting the shutter and the TTL signal, the shutter’s exposure can be controlled via a TTL 
square pulse input, the Front Panel pushbutton, or the USB COM port.   
Exposure time is determined by PULSE Input signal duration in the Low(0V) position (Active Low).  
This is referred to as Pulse Width Determined Exposure Time. 
The 

VLM1B

 can be easily integrated into OEM applications where a +5VDC power supply is available. 

Note that when a remote activate cable, the 710R (see 

https://www.uniblitz.com/products/710r-

handheld-remote-cable/

) or the 710R/F (see 

https://www.uniblitz.com/products/710rf-foot-activated-

remote-cable/

) can be plugged directly into the PULSE INPUT BNC only.  Depressing the switch on 

either cable will active the shutter to the open position. The Active state of this input must be Active 
Low.  If the state is active high, then +5V will be required to switch the shutter to the active state and 0V 

Содержание N-CAS VINCENT ASSOCIATES VLM1

Страница 1: ...User Manual VLM1 VLM1B Interface Module For the FS Shutter Series 14 0055 Version 1 01 2021 1 800 828 6972 ...

Страница 2: ... ongoing development program Vincent Associates reserves the right to discontinue or change specifications or designs at any time without incurring any obligation Revision Table Version 1 00 2020 10 21 20 STP Initial release Rev b Version 1 01 2021 3 23 21 STP Minor revisions Rev c Vincent Associates a Division of VA Inc 803 Linden Ave Rochester NY 14625 Tel 585 385 5930 Fax 585 385 6004 UNIBLITZ ...

Страница 3: ...onnections 9 VLM1 VLM1B Front Panel Operator Controls Indicators 10 Start Up 11 Extending the Shutter Interconnection 11 Terminal Block Wire Installation 12 OPERATING BASICS 14 Initial Operation and Testing 14 BNC I O Active State Selection 15 Fuse Information 15 USB Operation 18 Maintenance 18 Inspection 18 Cleaning Procedure 18 SPECIFICATIONS 19 System Characteristics 19 External Input Character...

Страница 4: ... Panel View Layout 9 Figure 3 VLM1 VLM1B Front Panel View Layout 10 Figure 4 VLM1 VLM1B SHUTTER Wire Installation Diagram 12 Figure 5 VLM1 to FS Series Shutter Interconnection Diagram 13 Figure 6 VLM1 Dimensional Layout 16 Figure 7 VLM1B Dimensional Layout Fuse Jumper Locations 17 Figure 8 VLM1 VLM1B Schematic Diagram 22 ...

Страница 5: ...product can be grounded through the SHUTTER connector Pin 4 if required Use Proper Fuse To avoid fire hazard use only the fuse type and rating specified for this product The VLM1 VLM1B contain no user replaceable fuses Do Not operate in Wet Damp Conditions To avoid electric shock do not operate this product in wet or damp conditions Do Not Operate in an Explosive Atmosphere To avoid injury or fire...

Страница 6: ...ations are described for all input output controls as well as operational features of the VLM1 VLM1B What follows is the complete operator s manual for the UNIBLITZ VLM1 VLM1B Interface Module Please read this manual completely before operating the unit Due to the construction of this unit Vincent Associates recommends that the unit be returned to the manufacturer for repair There are no user serv...

Страница 7: ...ires are easily inserted into the connector using a small 2mm screwdriver included Dimension a is 10mm See Figure 4 for the shutter wire installation procedure Figure 1 5 Pin Pluggable Phoenix Connector Exposure determined by external pulse width BNC 5V TTL active high active low selectable by board jumper Internal fuse protection for SHUTTER Connector Pin 1 and 5V Input USB 2 0 Micro B both are r...

Страница 8: ...he VLM1 VLM1B provides the switching circuit BNC interconnects USB interface etc to easily operate any shutter in the FS series Controlling the shutter from the front panel switch from any 5V TTL signal the duration of this pulse is approximately the exposure time of the shutter or a computer s USB port See Figure 8 which provides the Schematic Diagram for additional information on the VLM1 VLM1B ...

Страница 9: ...le with on board jumper See Figure 7 3 SYNC OUTPUT BNC For shutters equipped with electronic sync systems Provides a 5V TTL output signal Active High default or Active Low is selectable with on board jumper See Figure 7 when the shutter transfers to the active state 4 SHUTTER 5 Pin Pluggable Spring Terminal Connector For shutter termination and shutter sync system interface The flying leads from t...

Страница 10: ...s will indicate the shutter coil is Active and in the actuated state Please note that this LED and the SYNC Output will only activate if the shutter is equipped with the electronic sync and that the sync wires are connected properly to the unit 7 N C N O Manual Shutter Open Close Switch with GREEN LED Push the switch to open the shutter place in the active state for a Normally Open shutter the act...

Страница 11: ...me cable connect to the VLM1 or VLM1B 5 Plug in the power supply If being used to power the unit 6 Connect an input exposure signal 5V TTL to the input BNC if desired default Active Low 7 A push of the N C N O switch will cause the shutter to open The switch push button s GREEN LED will also illuminate indicating that a signal has been provided to the input switching MOSFET amplifier See VLM1 VLM1...

Страница 12: ... as shown 3 Slide wire into access hole below the terminal actuator 4 Release the terminal actuator 5 Wire should be connected a slight pull on the wire with the wire remaining stable and connected should indicate the wire is properly connected 6 Connect the remainder of the wires as required 7 For shutter operation RED wire from actuator coil must connect to Terminal 1 and BROWN wire from actuato...

Страница 13: ...VLM1 VLM1B User Manual 13 Figure 5 VLM1 to FS Series Shutter Interconnection Diagram ...

Страница 14: ...ols a MOSFET switch that in turn controls the FS Series Shutter from the 5VDC supply If the requirement is to operate the shutter from a switch contact closure follow this procedure this is when the active setting of the PULSE Input BNC is active low the default setting 1 Connect a wire from pin 4 of the SHUTTER connector this is the Ground connection Or short the outer shell of the BNC to the cen...

Страница 15: ... before attempting to change the jumpers Please read the entire manual carefully before attempting to operate the unit Fuse Information The VLM1 VLM1B are protected by resettable fuses on the 5V Input Power PWR COM connector and Shutter Actuator Coil A connection SHUTTER Pin 1 If a fault occurs when the power supply is connected to the unit or when the shutter was operating and then stops it could...

Страница 16: ...VLM1 VLM1B User Manual 16 Figure 6 VLM1 Dimensional Layout ...

Страница 17: ...VLM1 VLM1B User Manual 17 Figure 7 VLM1B Dimensional Layout Fuse Jumper Locations ...

Страница 18: ...ite 3 1 is copyrighted software that is free for personal and commercial use You may use it and distribute it without limitations You may however not remove or conceal the copyright There are no guarantees or warranties whatsoever use it at your own risk Maintenance Proper care and maintenance of the VLM1 VLM1B should be taken as with any sensitive electronic instrument There are no user serviceab...

Страница 19: ... PULSE INPUT BNC Active High Active Low input On board jumper selectable Default Active Low See BNC I O Active State Selection TTL Compatible Minimum High Level 2 0VDC Maximum low Level 0 8VDC Max Trigger Voltage on Input 10 0VDC Power Requirements PWR COM PWR COM Micro B Connector 5VDC 0 5A min USB5 1 0A wall adapter included with the VLM1 USB Requirements PWR COM USB 2 0 Driver should automatica...

Страница 20: ...EEN wire Sync Ground Pin 5 Shutter BLUE wire Sync 5VDC Fused with 125mA Fast Acting Fuse not serviceable General Characteristics Name Description POWER LED Indicator GREEN LED illuminates when power is connected SYNC Active Indicator YELLOW LED illuminates when the attached FS Shutter s Sync is active SYNC Output BNC will be active when the LED is illuminated FS Shutter used must have the Electron...

Страница 21: ...ft For Power or Computer Connection VLM1 Only Micro Screwdriver Flat Blade screwdriver for interconnection to 5 Pin spring terminal SHUTTER connection Accessories optional 710R Remote hand held trigger cable for connection to PULSE Input BNC Input must be set to active low 710R F Remote foot switch activated trigger cable to PULSE Input BNC Input must be set to active low 510FS 10 Ft 3m Extension ...

Страница 22: ...VLM1 VLM1B User Manual 22 Figure 8 VLM1 VLM1B Schematic Diagram ...