![Uni-Trend UTD2000E Series Скачать руководство пользователя страница 55](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099055.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
maximum limit 255 files.
The file name of internal storage bsv must be"wave01.bsv"~” wave255.bsv", maximum limit 255
files.
:FILE:SAVE
Command format:
:FILE:SAVE <filename>[,<source>][,<disk>]
Functional description:
For save channel waveform or setting data into file.
<filename> file name must be character string data with double quotation mark, for
example “test.bsv”
The file name of *.bsv or *.csv presents save channel waveform with the suffix name format
into file.
The file name of *.set presents save the setting data into file.
<source > presents physical channel
{CHANnel1 | CHANnel2| CHANnel3| CHANnel4}
,
optional parameter only valid when load waveform data.
CHANnel1 presents Channel 1
CHANnel2 presents Channel 2
CHANnel3 presents Channel 3
CHANnel4 presents Channel 4
<disk> presents memory medium {
FLASH | UDISK
}
,
optional parameter
,
ignore presents
internal data of FLASH.
FLASH presents internal data
UDISK presents U disk data
For example
:
FILE:SAVE "test.bsv",CHANnel1,UDISK
Save waveform data of channel 1 as test.bsv format into U disk.
FILE:SAVE "system-set-up01.set"
The configuration data of oscilloscope save as No.1 position of internal medium.
FILE:SAVE "wave01.bsv",CHANnel1,FLASH
Save waveform data of channel 1 into internal medium.
FILE:SAVE "wave01.bsv",CHANnel1
Save waveform data of channel 1 into internal medium.
FILE:SAVE "system-set-up01.set",FLASH
The configuration data of oscilloscope save as internal medium.
FILE:SAVE "system-set-up01.set"
The configuration data of oscilloscope save as No.1 position of internal medium.