![Uni-Trend UTD2000E Series Скачать руководство пользователя страница 15](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099015.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
:CHANnel<n>:OFFSet
Command format:
:CHANnel<n>:OFFSet <offset>
:CHANnel<n>:OFFSet?
Functional description:
Set wave offset in vertical position, <n> take value as 1/2/3/4/5, it presents {CH1/CH2/CH3/
CH4/ MATH }.
Return format:
Query return offset value by scientific notation method, unit is related with the current channel
unit.
For example:
:CHAN1:OFFS 20V
set channel 1 vertival offset as 20V
:CHAN1:OFFS?
query return 2.000e001
:CHANnel<n>:SCALe
Command format:
:CHANnel<n>:SCALe {<scale> | UP | DOWN}
:CHANnel<n>:SCALe?
Functional description:
Set voltage base in vertical position.
<scale>
:
voltage base value
;
UP
:
plus one position based on the current position
;
DOWN
:
minus one position based on the current position.
Return format:
Query return the current voltage position value by scientific notation method, unit is V.
For example:
:CHAN1:SCAL 20V
set channel 1 voltage base as 20V
:CHAN1:SCAL?
query return 2.000e001
。
:CHAN1:SCAL UP
plus one position based on 20V voltage base
:CHANnel<n>:UNITs
Command format:
:CHANnel<n>:UNITs {VOLTs|AMPeres }
:CHANnel<n>:UNITs?
Functional description:
Set channel unit as VOLTs and AMPeres.
Return format:
Query return VOLTs, AMPeres.