![Uni-Trend UTD2000E Series Скачать руководство пользователя страница 36](http://html1.mh-extra.com/html/uni-trend/utd2000e-series/utd2000e-series_manual_839099036.webp)
UNI-T
UTD2000E Series Osilloscope Programming Manual
Set the term of pulse time, GREaterthan, LESSthan and EQUal.
Return format:
Query return{ GREaterthan | LESSthan | EQUal }.
For example:
:TRIGger:PULSe:QUALifier GRE
set the pulse term as GREaterthan
:TRIGger:PULSe:QUALifier?
query return GREaterthan
:TRIGger:PULSe:POLarity
Command format:
:TRIGger:PULSe:POLarity {POSitive | NEGative}
:TRIGger:PULSe:POLarity?
Functional description:
Set pulse polarity, POSitive and NEGative.
Return format:
Query return{ POSitive | NEGative }.
For example:
:TRIGger:PULSe:POL POS
set pulse polarity as POSitive
:TRIGger:PULSe:POL?
query return POSitive
:TRIGger:PULSe:TIME
Command format:
:TRIGger:PULSe:TIME <time>
:TRIGger:PULSe:TIME?
Functional description:
Set the time interval of pulse width trigger.
Return format:
Query return
the current time interval, unit is s.
For example:
:TRIGger:PULSe:TIME 1
set the pulse width as 1s
:TRIGger:PULSe:TIME?
query return 1.000e000
:TRIGger:PULSe:TIME:UPPer
Command format:
:TRIGger:PULSe:TIME <time>
:TRIGger:PULSe:TIME?