Operation Steps:
1. Use the knob or arrow key to move the cursor to a specific line.
2. First select the modes of each step, and press
or the knob to change the current
mode. Stop when the desired mode appears, and switch the cursor to set the next
parameter by rotating the knob or pressing direction keys
3. The constant value setting only needs to input the number directly after the cursor is
selected, and then move the cursor to the next parameter that needs to be changed by
rotating the knob or pressing direction keys after completion.
4. Follow step 3 to set Time, Check, Min and Max. (Note: When the parameter you input
is not within the correct upper and lower limits, the cursor will not be able to move. You
need to use the delete button to clear the parameter or change it to the correct value.)
5. Other parameters can be modified in a similar way.
6. After setting, press the function key [Test] to enter the <List Mode> test page, as shown
in Figure 4-2-2-2.
Figure 4-2-2-2 Test Page of List Mode
7. Press the ON button, the electronic load starts to load, and the indicator light below the
button lights up. If you want to suspend the test, you can press the [Pause] button. If
you need to stop the load, press the ON button again, and the running indicator light
goes out.
8. After the test is completed, the user can press [Result] to view the test results. If the
test result is within the upper and lower limits set by the user, Pass will be displayed,
otherwise Fail will be displayed instead. The user can check whether each item is
passed and save the test record by taking a screenshot.
Note: The input data should not exceed the upper and lower limits of the current
parameter. When the data exceeds the parameter limits, they will not be saved, and the
cursor will stay in the grid. At this time, you need to clear or re-enter the appropriate range
of parameters. When you need to change the numbers in the list, click Delete to clear the
data in the table. After completing each parameter setting in the list, click the Save button
for easy calling next time.