UTL8500 Series User Manual
26
4. Select the storage location (Flash memory/U disk) before saving and press the save
button to finish saving; record the set group number so that it can be called next time.
<Step>
After setting the group number, rotate the knob to move the cursor to [Step] to set the
number of steps for the list test, up to 16 steps can be set.
<Repeat>
Rotate the knob to move the cursor to [Repeat] to set the number of repetitions for each
measurement in the current mode. You can set up to 99999 times.
<Mode>
Rotate the knob to move the cursor to [Mode] to set the current running mode. Err means
that the load will automatically stop loading if there is an abnormality during operation.
There are four options for setting the operating mode: Continues/Trig/Continues
+Err/Trig+Err.
Continues:
The load will continue to the next step after executing one step until the end
of the operation.
Trig:
The load will pause after executing a step, and wait for the trigger signal before
continuing to the next step.
Err:
The test will automatically stop when an overrun or other error occurs during the load
operation.
4.2.2.2 Parameter Setting of Mode
After setting Group, Step and Mode parameters, the mode parameters for the set step will
appear in the list below, as shown in Table 4-2-2-2. Modify the parameters according to the
required mode.
Table 4-2-2-2 Parameter Setting of Mode
List Value
Description
Mode
CC/CV/CR/CP/Open/Short
Load Operation Mode
Value
Set Constant Value
Open/Short Default Value Is 1
Time/ms
200~999999ms
Set the load execution time of
each step between 200~
999999ms.
Check
Off/Curr/Volt/Power
Select Check Item
Min
Minimum Value of the Check Item
Set Minimum Value of the
Check Item
Max
Maximum Value of the Check
Item
Set Maximum Value of the
Check Item