UTD4000 Four-channel User Manual
Steps :
A s i n t he p r e v io u s i l lu s t r a t i on , s e t t he
attenuation factor of the probe and Ch1.
Connect the signal to ensure stable display of
waveform.
Improving trigger by setting trigger coupling
MENU
R e d u c i n g d i s p l a y n o i s e b y s e t t i n g t h e
sampling mode.
ACQUIRE
F1
F3
1.
2.
See the previous illustration for
operation instruction. See the previous chapter
for guidance on adjusting the horizontal time
base and vertical range.
3.
Press [
] in the trigger zone to display the
trigger setup menu.
S e t
t r i g g e r
c o u p l i n g
t o
l o w
f r e q u e n c y
suppression
or high frequency suppression
By
selecting
low
frequency
suppression
a
high pass
filter
is
set
up
It
filters
low
frequency signal quantities under
kHz and
allows
high
frequency
signal
quantities
to
pass through
If you
select high frequency
suppression a low-pass filter is set up. It filters
high frequency signal quantities higher than
80kHz and allows low frequency signal quantities
to pass through. By setting low frequency
suppression
or
high frequency suppression
you
can
suppress
low
frequency
or
high
frequency noise respectively and achieve a
stable trigger.
4.
If the signal being measured is stacked with
random noise and the waveform is too coarse as
a result, you can use the average sampling mode
to eliminate random noise display and reduce the
size of waveform for easy observation and
measurement. After getting the mean, random
noise is reduced and details of the signal are
clearer. Follow the steps below :
Press [
] in the menu zone of the front
panel to display the sampling setup menu. Press
menu operation key [
] then [
] to set
sampling mode to AVERAGE, then press the
control knob to adjust the
average number of times in multiples of 2
.
.
-
.
8 0
.
,
,
①
②
MULTIPURPOSE