109
UTD4000 Four-channel User Manual
Vertical
Model
UTD4304C
A/D converter
8-bit resolution
Deflection graticule factor V/div Range
2mV/div ~ 5V/div at input BNC
Displacement range
±
div
5
Analog bandwidth
Low frequency response (AC coupling, -3dB)
UTD4204C
UTD4104C
300MHz
200MHz
100MHz
Single bandwidth
100MHz
100MHz
100MHz
Selectable analog bandwidth limit (Typical) 20MHz
≤
10Hz at BNC
Rise time
≤
1.2ns
≤
1.8ns
≤
3.5ns
DC gain accuracy
When vertical sensitivity is 2mV/div : 4% (sampling or average acquisition mode)
When vertical sensitivity is 5mV/div~5V/div : 3% (sampling or average acquisition mode)
±
±
DC measurement accuracy
(average acquisition mode)
When vertical position is zero and N 16 :
(5% x reading
0.1 div
1mV) and 2mV/div is selected
(3% x reading
0.1 div
1mV) and 5mV/div~5V/div is selected
When vertical position is not zero and N 16 :
3% x (reading
vertical shift reading)
(1% x vertical shift reading)
0.2div). Set from 5mV/div to 200mV/div plus 2mV
Setup value > 200mV/div to 5V/div plus 50mV
≥
≥
±
+
+
±
+
+
± [
+
+
] +
Voltage difference (
V)
measurement accuracy
(average acquisition mode)
△
Under identical setup and environmental conditions, the voltage difference
( V) between two points of the waveform after the average of 16 waveforms
acquired waveforms is taken :
(3% x reading
0.05div)
±
+
△
≥