Figure 6-13 Trigger – Slope Type (Page 2)
Table 6-7
Menu
Settings
Description
Type
Slope
Set trigger Type to Slope
Slope Type
Rise
Fall
Trigger on rising edge that traverse between thresholds.
Trigger on falling edge between thresholds
Qualifier
>
<
=
Trigger When the slope is greater than the set value of slew
rate.
Trigger When the slope is less than the set value of slew
rate.
Trigger When the slope equals to the set value of slew rate.
Next
----
Page 2 of Slope Menu
Slew Rate
----
To set up slew rate value, use Left and Right arrow buttons
to pulse width time when the menu is displayed.
Threshold
Low
High
Low&High
Use Up and Down buttons to move low-level threshold
Use Up and Down buttons to move High-level threshold
Use Up and Down buttons to move low-and high-level
thresholds.
Back
-----
To return to Page 1 of Slope Menu.
Term Explanation
1. Trigger Source:
Input channel is used as the trigger source
■
Input Channel:
Input channel (CHI) is the most commonly used trigger source. For the channel selected
as the trigger source, it should be switched on before it works normally.
2.
Trigger Mode:
decide what the oscilloscope will do when no trigger event occurs. Three trigger modes
are offered: Auto, Normal and Single.
■
Auto:
The system automatically acquires data even there is no signal input and therefore displays scan
reference line on the screen; when the trigger occurs, it converted into trigger scanning and synchronizes
with the signal on its own.
Note: The Oscilloscope will go into SCAN mode if waveform scan rate is set at 50ms/div or lower timebase.
■
Normal:
Under this mode, the Oscilloscope acquires the waveform only after the trigger conditions are
met. It stops to acquire and keeps waiting if no signal is input. Once the trigger occurs, the scan starts.
■
Single:
You just need to press RUN button if under single mode, then the Oscilloscope waits to trigger. It
begins to acquire and display all acquired waveforms once single-shot trigger signal is detected, after that,
it stops.
52
UTD1000L User Manual
Содержание UTD1000L Series
Страница 1: ...User Manual UTD1000L Series DIGITAL STORAGE OSCILLOSCOPE ...
Страница 2: ...2011 09 UNI T Technologies Inc ...
Страница 3: ......
Страница 60: ...Figure 6 17 Select Vpp Parameter Figure 6 18 Select Amplitude Parameter UTD1000L User Manual 57 ...
Страница 85: ...This user manual may be revised without prior notice 82 UTD1000L User Manual ...