Because the temperature transducer aluminum plate is put at the back of the
Calibrator, try to open the tilt stand if possible when operating the Calibrator.
Don’t touch the aluminum plate when operating the Calibrator to ensure the
accurate of compensation temperature.
If the operating temperature is changed, it is necessary to wait for around 10
minutes to stable the internal transducer before operating the Calibrator.
Follow the above 1 to 7 steps to remove the cold terminal temperature auto
compensation, but change the adjust digit to 1 on step 4.
Note:
Model UT713: OPERATING MANUAL
13