UNI-T
UPO 3000E&UPO2000E Series Oscilloscope Programming Manual
:SBUS:CAN:NEGative:SOURce
Command format:
:SBUS:CAN:NEGative:SOURce { CHANnel1 |
CHANnel2 |
CHANnel3 | CHANnel4 }
:SBUS:CAN:NEGative:SOURce?
Functional description:
Set input negative source of CAN bus
Return format:
Query return{ CHANnel1 |
CHANnel2 |
CHANnel3 | CHANnel4 }.
For example:
:SBUS:CAN:NEGative:SOURce CHANnel1
set channel 1 as CAN bus negative source
:SBUS:CAN:NEGative:SOURce?
query return CHANnel1
:SBUS:CAN:SIGNal:DEFinition
Command format:
:SBUS:CAN:SIGNal:DEFinition { CANH | CANL | RXTX | DIFFerential }
:SBUS:CAN:SIGNal:DEFinition?
Functional description:
Set signal mode of CAN bus.
Return format:
Query return{ CANH | CANL | RXTX | DIFFerential }.
For example:
:SBUS:CAN:SIGNal:DEFinition CANH
set CANH as the signal of CAN bus
:SBUS:CAN:SIGNal:DEFinition?
query return CANH
:SBUS:CAN:SAMPlepoint
Command format:
:SBUS:CAN:SAMPlepoint <percent>
:SBUS:CAN:SAMPlepoint?
Functional description:
Set the sampling percentage of CAN bus signal, <percent> range 1~99.
Return format:
Query return sampling percentage.
For example:
:SBUS:CAN:SAMPlepoint 30
sampling percentage as 30%
:SBUS:CAN:SAMPlepoint?
query return 30