11.5 SCR FUNCTION
1. Set the field meter to satellite frequency range (press and hold
) and in spectrum
analyzer mode (
).
Press and hold the
to activate the menu.
Select DiSEqC SCR and confirm with
.
Select MONO FEED and press
.
DiSEqC
2 WAY DiSEqC
4 WAY DiSEqC
DiSEqC Level
DiSEqC User
MOTOR (1.2)
DiSEqC SCR
MONO FEED
DUAL FEED
2. Select 1 of the 4 frequencies relative to the 4 SCR channels and press
.
DiSEqC SCR
SCR A : 1,210 MHz
SCR B : 1,420 MHz
SCR C : 1,680 MHz
SCR D : 2,040 MHz
SCR E : USER
3. After selecting one of the 4 channels to check, the instrument will return to the
SPECTRUM mode indicating SCR at the top right in order to signal the particular mode
of use.
4. To check correct operation of the selected SCR channel, you can use the pre-stored
transponders or else the parameters of the desired transponder must be manually set
as follows:
a. Set TRANSPODER frequency with the local oscillator and not 1IF frequencies,
for example TP124 RAI, type 10.992 and not 1.242,00.
Then press
and enter the numerical sequence followed by
.
b. Set the polarity through the side menu (
) POL V or POL H.