6. Installation
6.1. Installation site
○
Install STC-1C series in the site under proper usage environment.
CAUTION
Avoid using STC-1C series in these places.
・
Where flammable gas, corrosive gas, oil mist and particles that can
deteriorate electrical insulation are generated, or are abundant.
・
Where the temperature is below 0
º
C or above 40
º
C.
・
Where the relative humidity is 80%RH or below dew point.
・
Where highly intense vibration or impact is generated or transferred.
・
Near high voltage power lines or where inductive interference can affect the
operation of the product.
6.2. Installation
Please see
panel cut dimensions of figure 6-1 below, and machine the mounting
hole on your cabinet panel. Insert STC-1C series from the front panel into the
interior of the mounting hole completely, and fix it with two clamps.
Figure 6-1
Panel cut dimensions and installation method
93
93
K
Panel of User's Cabinet
(
Panel cut
)
Clamp holders mounting positions
(
Front
)
Tighten with a Phillips-head screwdriver.
(
Inside of user's cabinet
)
Содержание STC-1C Series
Страница 2: ......
Страница 8: ...F 6 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 12: ...Table of Contents IV ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 14: ...IW 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 40: ...CS 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...
Страница 42: ...Revision History R 2 ULVAC CRYOGENICS INCORPORATED This page intentionally left blank...