3.PREPARATION
8
■How to install and make holes in the rack
The RF power supply can be mounted on a rack. Refer to Fig. 3.3 below how to make mounting holes
in a rack.
Fig. 3.3 Rack installation hole position
■Structure for mounting RF power supply
In mounting the RF power supply in a front panel, see to it that undue force is not exerted to the
front panel of the RF power supply, which is not designed to bear the weight of the power supply.
See Fig. 3.4 below. Also provide a clearance for air intake around the power supply.
And, when using rack-mount, enable to ventilate enough in order that the temperature in a rack does
not exceed specification temperature. If it is operated at the temperature besides specification,
alarm may detect or it may damage in the worst. See Fig. 3.5 below.
RF Power supply
Rack
Rack
Spport from below
No undue force on
the front panel
power supply
Fig. 3.4 Mounting rack-mount type RF power supply (example)
100mm or more
40mm or more
50mm or
more
FAN
Top view of
a power supply
Front panel of
a power supply
Fig. 3.5 Clearance around power supply