ULVAC CRYOGENICS INCORPORATED
37
Problem
Possible Cause
Corrective Action
(9) LN
2
accumulation
declines
Ice or frost appears around the
cold head.
Contact us.
The maintenance interval for
cold head and oil absorber is
reached.
Perform the maintenance works,
depending on operation hours.
Helium gas pressure of
UW701N is decreased.
Fill the Helium gas additionally
until the specified pressure is
reached. When the trouble
occurs repeatedly, contact us.
Evaporation in the dewar
increases.
Contact us.
MP-300K inhales gases such
as helium.
Contact us.
The purity of nitrogen gas is
not sufficient.
Check the purity of the nitrogen
gas.
(10) AUTO READY button
flickers
The level sensor in the supply
destination is in error state.
Contact us.
The level sensor cable in the
supply destination is not
connected or broken.
Check the level sensor cable,
(11) MANUAL START button
flickers
30 seconds before the manual
dispense time limit (10
minutes) is reached.
If you wish to continue
dispensing, press MANUAL
START button again.
(12) AUTO READY and
AUTO SUPPLY flickers
alternately
Supply signal continues from
the level sensor in the supply
destination although automatic
transfer time limitation is
reached.
Press AUTO READY to release
the automatic supply state, and
check the LN2 level in the
supply destination.
Contact us if any error is found.
(13) Other failures
Contact us.