Documents are only to be used and distributed completely and unchanged. It is strictly the users’ responsibility to check carefully
the validity of this document with respect to his product.
page 8 of 24
No. 47800-2-02-5
Technical data
We reserve the right for technical modifications without prior notice!
Type
MD 4U NT
Maximum pumping speed 50/60 Hz (ISO 21360)
m
3
/h
3.8 / 4.3
Ultimate vacuum (absolute) without gas ballast
mbar
1
Ultimate vacuum (absolute) with gas ballast
mbar
3
Maximum permissible inlet pressure (absolute)
bar
1.1
Maximum permissible outlet pressure (absolute)
bar
1.1
Maximum permissible pressure (absolute) at
gas ballast valve
bar
1.2
Permissible ambient temperature
storage / operation
°C
-10 to +60 / +10 to +40
Permissible relative atmospheric moisture during
operation
(no condensation)
%
30 to 85
Rated motor power
kW
0.25
No-load speed 50/60 Hz
min
-1
1500 / 1800
Maximum permissible range of supply voltage
(+/- 10% ) dual voltage motor
Attention: Observe specifications on rating
plate!
100-115 V~ 50/60 Hz
120 V~ 60 Hz
200-230 V~ 50/60 Hz
Maximum rated current at:
100-115 V~ 50/60 Hz 120 V~ 60 Hz
200-230 V~ 50/60 Hz
A
A
5.7
3.0
Device fuse
slow blow fuse 6.3A
Motor protection
thermal cutout, manual reset
Degree of protection IEC 529
IP 40
Inlet
small flange KF 16
Outlet
silencer
Dimensions L x W x H approx.
mm
325 x 239 x 198
Weight approx.
kg
16.4