-24-
5.2 Maintenance
When maintenance interval has elapsed, be sure to contact our distributor who sold it
to you. This vacuum pump requires maintenance conducted only by our authorized
specialist.
Never try to disassemble, reassemble or alter on user’s side. We are not responsible
for any accidents caused by disassembly, reassembly or alteration which was done
by the user or non-specialist.
The following parts are consumable and need to be replaced periodically.
Whenever something goes wrong with them, replace them immediately.
W h e r e t o i n s p e c t
Maintenance interval
Yearly or
8,000 h
Biennially or
16,000 h
Triennially or
24,000 h
4th years or
32,000 h
Angular contact Ball bearing set
-
Grease /
△
-
○
Pin crank set
Grease /
△
Grease /
△
Grease /
△
○
Duplex arrangement angular ball
bearing set [Housing]
-
Grease /
△
-
○
Roller bearing set [OS]
Grease /
△
Grease /
△
Grease /
△
○
Spider
○
○
○
○
P-seal [FS set]
○
○
○
○
Tip seal FS
○
○
○
○
Tip seal OS
○
○
○
○
Air-flush kit
○
○
○
○
Cooling fan
△
△
△
○
Vacuum pump itself
-
-
-
○
○・・・
Replace
△・・・
Replace if something goes wrong.
Note 1
:
Be sure to use designated DISL exclusive grease.
Note 2
:
You must shorten maintenance standard when pumping vapor since vapor temperature ,
disposal volume, disposal frequency and substances in vapor have influence on pump
operation.
Note 3
:
The maintenance interval should be earlier one in either the period or running hours.
Note 2
:
When you want further
operation
after either the 4
th
year or 32,000 operating hours, please
contact our distributor who sold it to you.
Important
Causes of failure
Shorten maintenance interval if conditions of installation or operation are unfavorable.
In particular, ambient temperature has a great influence on failure. Maintenance interval is based on an ambient
temperature 5~40
℃
and a yearly average ambient temperature 25
℃
.
Shorten the maintenance interval if temperature exceeds the foregoing. If not, it can cause failure.
Maintenance interval is not a guarantee period.
Exceeding maintenance interval
Operation exceeding maintenance interval increases risk of failure and accidents.
When maintenance interval has elapsed, be sure to contact either our sales branches who sold it to you or local