06 

 

 
Installation 

(1)  The fine clearances used in this pump require that the following conditions be satisfied 

during storage, installation, and operation. 

 

1. 

Ambient temperature of 7~40°C and maximum relative humidity of 85% during 

operation. 

 

2.    Equal to or less than meters above the sea level 1000m storage and operation. 

 

3. 

Other conditions for storage and operation. 

 

 

a)  Level floor of sufficient strength. 

 

 

b)  No condensation 

 

 

c) 

Dust-free environment 

 

 

d)  Well ventilated 

 

 

e)  Environment free of corrosive or explosive gas. 

 

 

f) 

Not subject to direct sunlight. 

 

 

g)  No danger of fire. 

 

 

h)  Maximum ambient temperature of 40°C during assembly of pump. 

 

 

i)    Free of soot and oil. 

 

 

j)    Free of splashing or flooding. 

 

 

k)    Keep it indoor ventilated. 

 

l)    When installing the pump, avoid mounting a vacuum pump directly on the base. 

Employ anti-vibration rubber between the base and the vacuum pump.

 

 
(2)  To prevent back injury, always use both hands to lift pumps. 
(3)  Microscopic particles resulting from wear of components are discharged from the outlet and 

contaminate the room. If necessary, connect a pipe from the discharge outlet to the outside 
of the building. 

 
Operation 

(4)  Do not use in applications involving organ transplants, or contact with body fluids or living 

tissue. 

(5)  Touching rotating components (e.g. motor, main shaft, axial joints) while the pump is in 

operation may result in injury. 

(6)  The overload protector operates when the pump becomes excessively hot. Touching it in 

this condition may result in burns. 

(7)  Touching the motor while the pump is in operation or while it is still hot immediately after 

having been switched off may result in burns. 

(8)  Do not insert fingers or objects into, or peer into, the inlet or outlet during operation. 
(9) 

 

Ensure that the customer installs a trap if steam is generated. Ensure that condensate does 
not enter the vacuum pump. A malfunction may result if condensate enters the pump.

 

 

Maintenance and Repair 

(10)   Dispose in accordance with legislation for disposal and cleaning of waste products, handle 

as industrial waste, and do not incinerate. 

(11)  If the pump ceases operation, turn power OFF immediately to prevent accidents, remove 

the power cord from the terminal, and contact your dealer or the manufacturer for 
inspection and repair. 

(12)  Leave the pump for at least 30 minutes until it has cooled, and begin operation again. 

Touching the pump immediately after it has stopped may result in burns. 

(13) As the pump is not protected against entry of water, it is not guaranteed against splashing 

or flooding. 

 

 

 

!

Caution

 

Содержание DAT-50DA

Страница 1: ...ior to use For safe and efficient use of this pump please read this manual carefully before operation After reading the manual keep it in your file for future reference Specifications in this manual a...

Страница 2: ...umps Safety requirements Part2 Vacuum pumps IEC EN 61010 1 2010 A1 2019 Safety requirement for electrical equipment for measurement control and laboratory use Part1 General requirement following the p...

Страница 3: ...th the following standards BS EN 1012 2 1996 A1 2009 Compressors and vacuum pumps Safety requirements Part2 Vacuum pumps BS IEC EN 61010 1 2010 A1 2019 Safety requirement for electrical equipment for...

Страница 4: ...s in the power voltage and frequency 4 3 6 Checking Operation After Installation 5 3 7 Piping 5 3 8 Storage 5 4 Cautions for Operation 5 4 1 Cautions for Operation 5 4 2 Operation of the Thermal Prote...

Страница 5: ...hange region of the voltage and frequency 4 Fig 3 2 Example of Piping Used When Evacuating a Vessel 5 Fig 6 1 Disassembly drawing of consumable 10 Table 1 Product Specifications 1 Table 2 Consumables...

Страница 6: ...quipment correctly Read the section on Safe Use particularly closely Keep This Manual in a Safe Place After reading this manual be sure to keep it in a safe place which is readily accessible to others...

Страница 7: ...e sales division of the manufacturer if there are any problems with the pump 1 Ensure that the tube on the side of the pump is not gripped or bent when removing the pump from the packaging Holding the...

Страница 8: ...ct handling of the equipment may result in death or serious injury to the operator Incorrect handling of the equipment may result in light or medium injuries to the operator or damage to the equipment...

Страница 9: ...materials such as solvents in the vicinity when using the pump 3 Ensure that the motor is freely ventilated to prevent overheating which may result in fire or burns Power Supply 4 Always remove the po...

Страница 10: ...result in electric shock 14 The electric motor fitted to this pump incorporates a thermal protector however it is not guaranteed to be open when it eventually fails at some time For safety reasons th...

Страница 11: ...ons involving organ transplants or contact with body fluids or living tissue 5 Touching rotating components e g motor main shaft axial joints while the pump is in operation may result in injury 6 The...

Страница 12: ...e pump with a pressurized air supply 7 Ingestion of liquids into the pump will result in damage and prevent proper operation Ingestion of rubbish and dust in the air entering the pump will interfere w...

Страница 13: ...at near atmospheric pressure 1 2 Specifications Table 1 Product Specifications Model DAT 50DA DAT 100SA Pumping Speed L min 50Hz 50 100 60Hz 55 110 Ultimate Pressure kPa 3 3 15 0 Motor 3 200 220V 150W...

Страница 14: ...2 2 Dimensions Fig 2 2 DAT 100SA Dimensions Fig 2 1 DAT 50DA Dimensions...

Страница 15: ...n the base Employ anti vibration rubber between the base and the vacuum pump 3 3 Location The pump should be installed level in a location with minimal dust and humidity This location should be select...

Страница 16: ...the voltage change and frequency change in Domain B it shall operate with main rated values and shall be used practically convenient However operation with it is convenient and safe is maintained on p...

Страница 17: ...ipe and the vessel see Fig 3 2 Fig 3 2 Example of Piping Used When Evacuating a Vessel 3 8 Storage Turn the power off remove the terminal place the rubber caps over the inlet and outlet and store the...

Страница 18: ...Commence normal operation once the pump has warmed 5 Pump Performance 5 1 Pressure Achieved The term pressure achieved as employed in the catalogue and in this manual is defined as the minimum pressur...

Страница 19: ...ctions Inspect consumables after the first 6000 hours of operation and replace and clean in accordance with the Replacement and Cleaning Guide on the following page Refer to 6 4 Replacing and Cleaning...

Страница 20: ...ration After stopping the pump leave it for 30 minutes to cool and replace and clean components only after it has cooled to a safe temperature Always put on dust mask and gloves before replacing diaph...

Страница 21: ...placing head gaskets Remove the five cross recessed head machine screws M4 12 and remove the head cover Open piece each of black head gasket is located in the head cover for one head total two Remove...

Страница 22: ...10 DAT 100S DAT 50D Fig 6 1 Disassembly drawing of consumable...

Страница 23: ...vessel 1 Select another pump 2 Pressure measurement is incorrect 2 Measure the pressure correctly 5 1 Pressure does not diminish 3 Vacuum gauge is unsuitable 3 Measure with a calibrated vacuum gauge s...

Страница 24: ...e and inspection etc d Malfunctions caused by modifications or repairs carried out by a party other than the manufacturer or by a service company not approved by the manufacturer e Malfunctions caused...

Страница 25: ...Industrial Safety and Health Law designates particular substances as the materials to be notified 2 Usage Status Operation Method Approx hours per day years and months Continuous Operation Intermitte...

Страница 26: ...6 6397 2286 ULVAC KIKO Inc https ulvac kiko com en Please contact us for products Service Base or other Inquiries from here ULVAC Inc Components Division 2500 Hagisono Chigasaki Kanagawa 253 8543 Japa...

Отзывы: