83
NT UF 811 CP GB (e) / June 2015
Presence of noise (acoustic interference)
Comment:
For the last three cases, a weak signal (gain greater than 50 dB) may mean that the
probes require realignment or maintenance (pollution on the surface of the probe).
Presence of electromagnetic interference
Comment:
in this case, move the cables (in particular the probe cables) away from the power
cables. Check the ground connections and the shielding.