17
NT UF 811 CP GB (e) / June 2015
3.3.1 Communication connector
Depending on whether the RS232 or RS485 interface has been selected, connect the RX TX or A, B pins to
ground.
Notes:
The serial link is electrically insulated.
If the UF 811 is the last device on the RS485 network, a termination resistance of 120 Ohm
must be added to the connector between pins A and B.
Only one of the two interfaces may be in operation at any one time.
The connector used is a
male 5-pin M12 type
connector with inverted B coding.
View of the base
The pins have the following functions:
Pin
Functions
1
RS 485 : A / RTx (-)
2
RS 485 : B / RTx (+)
3
RS 232 : Tx (Output Uf 811)
4
RS 232 : Rx (Input Uf 811)
5
GND ISO
Note
: the pin number is engraved directly on the connector.