- 18 -
7.
Turn the rotation switch to ON and dial speed to 1. Press START. Stop at the first (right) tilt knob
location. If the readout on the micrometer is not at ‘100’ adjust the tilt knob to bring the micrometer
needle to ‘100.’
8.
Rotate the head again to the second (left) tilt knob location. If the readout on the micrometer is not
at ‘100’ adjust the tilt knob to bring the micrometer needle to ‘100.’
9.
Rotate back to the pivot point. If the readout on the micrometer is not at ‘100,’ turn the Z-micro
adjust dial on top of the head to bring the head down so that the deflection on the Micrometer reads
‘100.
10.
Continue steps 7-9; adjusting the corresponding tilt knobs until there is no difference in height
between the three positions.
Note: IF ULTRACOLLIMATOR is available, its use (detailed in Section 5 below) offers much higher accuracy for
many applications.
Pivot Point
Left Tilt Knob
Right Tilt Knob