VLS Service Manual - REV2015.04
Adjustments and Settings
Page | 15
Rotary Calibration
1. Turn ON the UCP and the VLS.
2. Open the Top Door.
3. Lower the table about halfway down using the arrows on the UCP or the laser system keypad.
4. Turn OFF the VLS.
5. Properly mount the Rotary on top of the engraving table, making sure that it is seated properly in the
bracket.
NOTE:
If this is your first time installing the Rotary fixture. Please refer to the User Guide to install the Rotary
mounting bracket.
6. Turn ON the VLS.
7. The rotary automatically rotates to indicate it is properly connected.
8. Select the
System
tab in the UCP. In the Rotary field click CALIBRATE.
Adjustments and Settings
9. The Rotary Calibration window will appear with Y Position and Z Position boxes. In the Y Position box
uses the Y-axis buttons to move the focus carriage back and forth. Place the focus carriage at exactly
2.625”.
10. Adjust the X-axis using the buttons to move the focus carriage left and right and place the red LED over
the flat part of the concave metal fixture. This is normally located on the left side when the Rotary is
installed.
11. Adjust the Z-axis using the buttons or the Z-axis Menu on the keypad (recommended) to move the table
up and down.