6.11. Operating the manoeuvring aid
The treatment chair is equipped with 3 lowerable fixed castors to manoeuvre it to its final location. If necessary,
these can be turned off by stepping the foot pedal (see illustration) completely down. The treatment chair is
then mobile and easily manoeuvrable thanks to the integrated swivel castor.
2 fixed castors (rigid) and 1 swivel castor raised/retracted
Treatment chair stands stable on 4 adjustable feet
2 fixed castors (rigid) and 1 swivel castor extended
Treatment chair mobile
Caution
•
The manoeuvring aid may only be used for changing the location of the treatment chair (without the
patient!) or for the purpose of cleaning the room. Disconnect the mains connection cable from the
socket before moving the chair!
•
The treatment chair must not be loaded when it is mobile (manoeuvring aid activated).
•
Make sure that the patient can only enter the chair when it is safe locked (fixed castors retracted).