QUICK SET UP
PIPING
Install in pipe making sure to orient IN port to flow supply.
10 pipe diameters distance is required upstream and 5 down for best accuracy.
Use proportional spacing if this much space is not available.
50 pipe diameters are required upstream as distance from a valve.
No use of Teflon tape please. (See detailed piping instructions.)
CONFIGURATION OF METER
At start up , display shows 1.888.8.8 showing that its digits are all functional
18888.8.8
Then model is shown
CP6 1.0
Example
Run mode achieved. Flow rate displayed at 0 flow is 0.
0.00
Note that if no adjustments are made, GPM displays.
SELECTING GPM, LPM OR CMH
Depress MENU button to display flow in GPM, LPM or CMH.
TOTALIZER FUNCTION
Depress SELECT button to display total flow. To change between total in GPM, LPM or CMH,
depress MENU button after entering totalizer mode.
Resetting total: Depress and hold both MENU and SELECT buttons to reset the totalizer. TOTAL
will blink 3 times.