uide

 

.

7

Rotate test selector to Exch 
Test. Call for heat on the 
system. Observe and wait 
for O2 readings to stabilize.

Prior to the blower turning 
on, and after readings have 
stabilized, press the SEND 
button.  This will store the 
Pre-Blower test segment.

After the blower turns on 
press PUMP to start the Post- 
Blower test. 

The meter will wait 30 sec-

onds and then record the 

Post-Blower values for CO, O2 

and Excess Air.  Test results 

will automatically be stored  

to exchange reports. Report 

includes both Pre and Post 

Blower test segments.

To view test results rotate test 

selector to “MENU” go to the 

“REPORT” screen and select 

EXCH.  Select “VIEW” and 

scroll to desired log. Press 

“SEND” to print.

Rotate test selector to 
“MENU” 

scroll

 

s

UP or

 

t

DOWN to 

SCREEN. 

Press “SEND”.

Scroll

 

s

UP or

 “

t

DOWN 

to select the LINE you wish to 
change

Press “SEND”. 

(

LINE 1 shown)

Scroll

 

s

UP or

 “

t

DOWN 

to select the function you 
wish to display

Press 

“SEND”. 

(

Pressure shown)

Rotate selector to “AUX” 
to view. Pressure is now 
shown in line 1 along with 
T1, T2 and Differential/ 
delta T.

Scroll

 

s

UP or

 “

t

DOWN 

to 

AUX. 

Press “SEND”.

h

EAt

 E

XChAnGEr

 t

Est

 

 

There are many methods to test heat exchanger integrity.  One of these is to observe the Excess Air, O2 and CO readings both before and after the 

blower turns on.  If the heat exchanger is sealed your O2 and CO readings should remain fairly stable.  A breach in the heat exchanger may allow fresh air to 
be forced into the flue after the blower turns on due to a pressure increase in the plenum.  The result may  be a rise in the measured O2 in the stack gas and an 
increase in the Excess Air .  In some sealed systems the fresh air drawn in through the breach may reduce the combustion air available leading to an increase in 
the CO reading.  If either of these situations are present it is probable that there is a problem with the Heat Exchanger  which may require additional testing and 
inspection .

NOTE: Many cracks are invisible to borescopes or the naked eye, and only open or separate from pressure or temperature changes during operations.

No probes or hose connec-

tions required for this test. 

Place handset in the area to 

be tested. Rotate test selector 

to Room CO

Press the PUMP button to start 

the test. CO readings  will be 

logged every 2 minute for a 30 

minute time span

To view results rotate test 

selector to “MENU” and 

select “REPORT” Scroll to 

“ROOM CO” and press 

“SEND”. 

Press “SEND” to VIEW 

results.

CO r

OOm

 t

Est

AUX s

EttinG

Press “SEND” again to print.

Great for checking for ambient CO and back drafting situations.

Programmable Auxiliary screen allows for Tech selectable test parameters to be chosen.

Содержание EagleX

Страница 1: ...1 Eagle Eagle X Owners Manual...

Страница 2: ...Charge Indicator Light 4 Line Backlit Display Display Line Lights Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Flue Probe Temperature Plug Plugs into T1 Narr...

Страница 3: ...ulated heat lost turning the Carbon in the fuel to Carbon Dioxide CO2 Wet Calculated heat lost turning the Hydrogen in the fuel into water H2O CO Loss Calculated loss due to partially burnt Carbon Any...

Страница 4: ...as soon as possible Use the GAS ZERO function Eagle X only to purge the analyzer of excess gases To do this on a standard Eagle remove the probe from the flue and turn ON the pump Always allow the rea...

Страница 5: ...ng testing Connect flue probe thermocouple connector to T1 and connect flue probe to water trap as shown above Use optional probe with T2 for inlet temperature Insert Flue Probe in stack Adjust the co...

Страница 6: ...raft probe for measuring pressure Press and hold t DOWN button to zero pressure sensor Rotate test selector to Temp Connect flue probe thermocouple or accessory thermocouple connector to T1 Connect ac...

Страница 7: ...ss Air O2 and CO readings both before and after the blower turns on If the heat exchanger is sealed your O2 and CO readings should remain fairly stable A breach in the heat exchanger may allow fresh a...

Страница 8: ...inch mmHg Millimeters of mercury hPa Hecto pascals EXIT Return to main menu REPORT COMB N VIEW When viewing reports PRESSURE DEL ALL Press s UP or t DOWN to select report EXCH Press and hold s UP or t...

Страница 9: ...04 to 06wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Smoke 0 or manufacturer s recommendation Positive Overfire Gas Oil Oxygen 3 to 9 O2 Stack Draft Water Colum...

Страница 10: ...Verify proper combustion O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Send and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas...

Страница 11: ...mperature Rise AC side Static Pressure Drop across coils Oil Furnace Verify proper combustion O2 CO Stack Temp Stack Draft Efficiency Test Verify Smoke Set Up Over Fire Draft Verify proper Static Duct...

Страница 12: ...ce between the ambient pressure level and the pressure level in the flue This is created either by the natural buoyancy of the hot gases created in combustion lifting or by an inducer fan that assists...

Страница 13: ...product of combustion and from wet fuel is recovered and the wet loss term is zero Gross efficiency calculations assume that the energy contained in the water vapor is not recovered Since the fuel air...

Страница 14: ...vents or strong detergents as they may damage the finish impair safety or effect the reliability of the structural components Emptying Cleaning the In Line Water Trap The in line water trap should be...

Страница 15: ...ations Electromagnetic Compatibility EMC This product has been tested for compliance with the following generic standards EN 50081 1 EN 50082 1 and is certified to be compliant The European Council Di...

Страница 16: ...rights You may also have other rights which vary from state to state U S A 1 8 0 0 5 4 7 5 7 4 0 F a x 5 0 3 6 4 3 6 3 2 2 C A N A D A 1 8 7 7 4 7 5 0 6 4 8 F a x 6 0 4 2 7 8 8 2 9 9 W W W U E i T E S...

Отзывы: