background image

14

G

EnErAL

 m

AintEnAnCE

 

 

•  Calibrate your instrument annually to ensure it meets original performance specifications

 

•  Keep your instrument dry. If it gets wet, wipe dry immediately. Liquids can degrade electronic circuits

 

•  Whenever practical, keep the instrument away from dust and dirt that can cause premature wear

 

•  Although your instrument is built to withstand the rigors of daily use, it can be damaged by severe  

 

  impacts. Use reasonable caution when using and storing the meter

p

EriODiC

 s

ErviCE

 

WARNING!

 

Repair and service of this instrument is to be performed by qualified personnel only. Improper repair or  

 

service could result in physical degradation of the instrument. This could alter the protection from personal  

 

injury this meter provides to the operator. Perform only those maintenance tasks that you are qualified to do.

A

nnUAL

 r

E

-C

ALibrAtiOn

 

 

While the sensor has an expected life of more than two years in normal use it is recommended that the  

 

analyzer is re-calibrated at least annually, This is so that long-term drift on the sensor and electronics can be  

 

eliminated. Local regulations may require more frequent re-calibration and users should check with  

 

appropriate authorities to ensure they comply with relevant guidelines.

C

LEAninG

 

Periodically clean your instruments case using a damp cloth. DO NOT use abrasive, flammable liquids,  

 

cleaning solvents, or strong detergents as they may damage the finish, impair safety, or effect the reliability  

 

of the structural components.

E

mptyinG

 & C

LEAninG

 

thE

 i

n

-L

inE

 w

AtEr

 t

rAp

 

 

The in-line water trap should be checked and emptied on a regular basis. Water vapor will condense in the 

 

probe line, which may cause the water trap to fill suddenly if the probe is moved. Care should be taken  

 

at all times.

 

 

Carefully remove the rubber plug from the bottom 

 

of the water-trap housing. Dispose of the condensate in 

 

a suitable drain, care must be taken as it could be acidic.

 

If condensate spills onto the skin or clothing, clean off 

 

immediately using fresh water, seek medical advice 

 

if problems occur. Ensure plug is replaced before  

 

performing combustion tests. Note: O2 reading will  

 

be high if the Water Trap Plug is not in place.

C

hAnGinG

 t

hE

 p

ArtiCLE

 f

iLtEr

 

This is a very important part of the analyzer and should be changed regularly. It prevents dust and dirt  

 

particles from entering the pump and sensors that will cause damage. The filter MUST be changed when 

 

it appears discolored on the inner surface.

 

Remove water-trap assembly from the analyzer  

 

as shown above. Remove the filter and plastic 

 

holder from the housing. Discard the filter  

 

element but keep the holder to fit to the new  

 

filter. Clean the inside of the filter housing  

 

with a suitable soft cloth. Fit the holder onto  

 

the new filter element and then insert into the  

 

housing. Refit the housing onto the analyzer.

Hook slides

out of dock

Slide outward

Pull plug out 

to drain

Insert a new filter

G

EnErAL 

m

AintEnAnCE

Содержание EagleX

Страница 1: ...1 Eagle Eagle X Owners Manual...

Страница 2: ...Charge Indicator Light 4 Line Backlit Display Display Line Lights Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Flue Probe Temperature Plug Plugs into T1 Narr...

Страница 3: ...ulated heat lost turning the Carbon in the fuel to Carbon Dioxide CO2 Wet Calculated heat lost turning the Hydrogen in the fuel into water H2O CO Loss Calculated loss due to partially burnt Carbon Any...

Страница 4: ...as soon as possible Use the GAS ZERO function Eagle X only to purge the analyzer of excess gases To do this on a standard Eagle remove the probe from the flue and turn ON the pump Always allow the rea...

Страница 5: ...ng testing Connect flue probe thermocouple connector to T1 and connect flue probe to water trap as shown above Use optional probe with T2 for inlet temperature Insert Flue Probe in stack Adjust the co...

Страница 6: ...raft probe for measuring pressure Press and hold t DOWN button to zero pressure sensor Rotate test selector to Temp Connect flue probe thermocouple or accessory thermocouple connector to T1 Connect ac...

Страница 7: ...ss Air O2 and CO readings both before and after the blower turns on If the heat exchanger is sealed your O2 and CO readings should remain fairly stable A breach in the heat exchanger may allow fresh a...

Страница 8: ...inch mmHg Millimeters of mercury hPa Hecto pascals EXIT Return to main menu REPORT COMB N VIEW When viewing reports PRESSURE DEL ALL Press s UP or t DOWN to select report EXCH Press and hold s UP or t...

Страница 9: ...04 to 06wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Smoke 0 or manufacturer s recommendation Positive Overfire Gas Oil Oxygen 3 to 9 O2 Stack Draft Water Colum...

Страница 10: ...Verify proper combustion O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Send and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas...

Страница 11: ...mperature Rise AC side Static Pressure Drop across coils Oil Furnace Verify proper combustion O2 CO Stack Temp Stack Draft Efficiency Test Verify Smoke Set Up Over Fire Draft Verify proper Static Duct...

Страница 12: ...ce between the ambient pressure level and the pressure level in the flue This is created either by the natural buoyancy of the hot gases created in combustion lifting or by an inducer fan that assists...

Страница 13: ...product of combustion and from wet fuel is recovered and the wet loss term is zero Gross efficiency calculations assume that the energy contained in the water vapor is not recovered Since the fuel air...

Страница 14: ...vents or strong detergents as they may damage the finish impair safety or effect the reliability of the structural components Emptying Cleaning the In Line Water Trap The in line water trap should be...

Страница 15: ...ations Electromagnetic Compatibility EMC This product has been tested for compliance with the following generic standards EN 50081 1 EN 50082 1 and is certified to be compliant The European Council Di...

Страница 16: ...rights You may also have other rights which vary from state to state U S A 1 8 0 0 5 4 7 5 7 4 0 F a x 5 0 3 6 4 3 6 3 2 2 C A N A D A 1 8 7 7 4 7 5 0 6 4 8 F a x 6 0 4 2 7 8 8 2 9 9 W W W U E i T E S...

Отзывы: