14
G
EnErAL
m
AintEnAnCE
• Calibrate your instrument annually to ensure it meets original performance specifications
• Keep your instrument dry. If it gets wet, wipe dry immediately. Liquids can degrade electronic circuits
• Whenever practical, keep the instrument away from dust and dirt that can cause premature wear
• Although your instrument is built to withstand the rigors of daily use, it can be damaged by severe
impacts. Use reasonable caution when using and storing the meter
p
EriODiC
s
ErviCE
WARNING!
Repair and service of this instrument is to be performed by qualified personnel only. Improper repair or
service could result in physical degradation of the instrument. This could alter the protection from personal
injury this meter provides to the operator. Perform only those maintenance tasks that you are qualified to do.
A
nnUAL
r
E
-C
ALibrAtiOn
While the sensor has an expected life of more than two years in normal use it is recommended that the
analyzer is re-calibrated at least annually, This is so that long-term drift on the sensor and electronics can be
eliminated. Local regulations may require more frequent re-calibration and users should check with
appropriate authorities to ensure they comply with relevant guidelines.
C
LEAninG
Periodically clean your instruments case using a damp cloth. DO NOT use abrasive, flammable liquids,
cleaning solvents, or strong detergents as they may damage the finish, impair safety, or effect the reliability
of the structural components.
E
mptyinG
& C
LEAninG
thE
i
n
-L
inE
w
AtEr
t
rAp
The in-line water trap should be checked and emptied on a regular basis. Water vapor will condense in the
probe line, which may cause the water trap to fill suddenly if the probe is moved. Care should be taken
at all times.
Carefully remove the rubber plug from the bottom
of the water-trap housing. Dispose of the condensate in
a suitable drain, care must be taken as it could be acidic.
If condensate spills onto the skin or clothing, clean off
immediately using fresh water, seek medical advice
if problems occur. Ensure plug is replaced before
performing combustion tests. Note: O2 reading will
be high if the Water Trap Plug is not in place.
C
hAnGinG
t
hE
p
ArtiCLE
f
iLtEr
This is a very important part of the analyzer and should be changed regularly. It prevents dust and dirt
particles from entering the pump and sensors that will cause damage. The filter MUST be changed when
it appears discolored on the inner surface.
Remove water-trap assembly from the analyzer
as shown above. Remove the filter and plastic
holder from the housing. Discard the filter
element but keep the holder to fit to the new
filter. Clean the inside of the filter housing
with a suitable soft cloth. Fit the holder onto
the new filter element and then insert into the
housing. Refit the housing onto the analyzer.
Hook slides
out of dock
Slide outward
Pull plug out
to drain
Insert a new filter
G
EnErAL
m
AintEnAnCE
Содержание EagleX
Страница 1: ...1 Eagle Eagle X Owners Manual...