english
english
EN-54
EN-55
Technical information
Technical information
E
E
3. Customer service
The unit may only be repaired by the manufacturer or by an expressly au-
thorised organisation. Please contact:
UEBE Medical GmbH
Bürgermeister-Kuhn-Straße 22
97900 Külsheim, Germany
4. Original spare parts and accessories
The following original spare parts and accessories are available from spe-
cialist dealers:
Subject to technical modifications.
• Cuff type H6 (for wrist circumference 13.5-21.5 cm)
Part. no. 2106001, PZN 11145115
• Charger type L1 with USB charging cable
Part. no. 2106020, PZN 11145121
5. Technical Inspection / calibration check
Basically, a metrological inspection is recommended at intervals of 2
years. However, professional users in Germany are obligated to comply
with the aforementioned according to „Regulation for Operators of Med-
ical Devices”.
This can be implemented either by UEBE Medical GmbH, an authority
responsible for metrology, or authorised maintenance services. For this,
please observe your national provisions.
Upon request, responsible authorities or authorised maintenance services
receive a „Test instruction for metrological inspection“ from the manu-
facturer.
Please include the manual when delivering the monitor for technical in-
spection.
Important: No modifications, e.g. opening the unit may be made to this
unit without the manufacturer’s prior permission.
6. Technical data
Power supply:
Built-in lithium-ion battery, 3.7 V, 420 mAh, 5 VDC, 1 A,
service life: 300 charge cycles
Charger:
Type-L1 USB charger, model BLJ06L050100U-V
Input 100–240 VAC, 50–60 Hz, max. 0.2 A, output
5 VDC, 1 A
Display:
Digital LCD display, visible field 46.5 mm x 36.5 mm
Measurement procedure: Oscillometric measurement
Reference procedure of
clinical testing:
Auscultatory measurement
Pressure display range:
0 – 299 mmHg
Measurement range:
Systolic: 60 – 230 mmHg
Diastolic: 40 – 130 mmHg
Pulse: 40 – 199 pulses/min
Measurement precision
at operation conditions:
Pressure measurement: ± 3 mmHg
Pulse measurement: ± 5 %
Operating conditions:
Ambient temperature 5 to 40 °C, rel. air humidity 15 to
90 %, air pressure 700 to 1060 hPa
Storage and transport
conditions:
Ambient temperature -20 to 60 °C, rel. air humidity up
to 93 %, not condensing
Time to adjust from min-
imum/maximum storage
temperature to operating
temperature:
30 minutes
Cuff:
For wrist circumference 13.5-21.5 cm
Weight:
approx. 103 g
Dimensions:
approx. 80 mm x 73 mm x 13 mm
Accessories:
Charger, USB charging cable, user manual, blood pres-
sure diary, storage bag
Mode of operation:
Continuous operation
Type of the application
part:
BF
Maximum temperature of
the application part:
42.5°C at an ambient temperature of 40°C
Classification:
Internal power supply
Charger: protection class II
IP Classification:
IP 22: Protection against solid foreign bodies with a
diameter of more than 12.5 mm; protection against
small amounts of liquid.
[email protected]