MENU 05 DISPLAY MODE

 

THERE ARE THREE MODES TO CHOOSE IN DISPLAY MODE: REAL TIME, SNAPSHOT 
AND PEAK HOLD  

Real Time

 is the standard operation of the instrument. For basic inspection operations choose 

Real Time. Snapshot is a very useful mode for inspections that require a comparison of 
measurements.  

Snapshot

 holds a specific reading on the display. The display can be updated by releasing and 

pressing the trigger. An example of this operation mode would be to locate the loudest point in 
a machine. By pointing the instrument at a loud signal and pressing the trigger, the sound 
intensity level will be displayed on the panel and held for comparison as the instrument is 
scanned around other points on the machine. The meter will remain constant while the audio 
levels change.  Another example is performing a quick comparison of multiple bearings by 
pressing and releasing the trigger to update and compare sound levels.   

Peak Hold

 displays and holds the peak value for comparison. It changes only when a higher 

ultrasound level is sensed. The bar graph will move up and down to display sound intensities, 
but the Peak Hold dB reading in the upper left corner will remain constant. A thin vertical line 
on the bar graph indicates the peak intensity of the bar graph. The Peak Hold dB reading is 
reset by turning the instrument off or by changing the frequency. 

TO SELECT DISPLAY MODE 

 

Make sure the Ultraprobe is off. 

 

Press (click) both the yellow Store button and the Sensitivity dial at the same time, then 
squeeze and hold the trigger. 

 

When in the first Menu Selection: “Data Transfer” (Menu 01), th

e user may move to any of 

the other Menu Selections by spinning the Sensitivity Control up or down (clockwise or 
counter clockwise). 

 

Spin to “Display Mode” (Menu 05) blinks.

 

 

Click the Sensitivity Control to enter Display Mode. 

 

Spin the Sensitivity Control dial until the desired setting (Real Time, Snap Shot or Peak 
Hold) appears and blinks. 

 

Click the Sensitivity Control Dial to set and return to Set Up Mode. 

 

S

pin to “Exit to PGM” Exit to Program (Menu 10) blinks. Click to enter Operation Mode

   

 

Содержание Ultraprobe 9000

Страница 1: ...0 UE SYSTEMS INC THE ULTRAPROBE 9000 MANUAL...

Страница 2: ...f Switch I O Port Battery Compartment Battery Wrist Strap Sensitivity Control Dial Storage Entry Button Head Set Jack Recharge Jack Standard Accessories Headset WTG 1 Warble Tone Generator Rubber Focu...

Страница 3: ...ext Editor Data Transfer Set Time Date dB Scale Select dB Offset Val Value This is a Reference dB Level Display Mode Calibration Due Date Text Editor Date Format Factory Defaults Exit to Program Triso...

Страница 4: ...ARE IMPORTANT TO UNDERSTAND OPERATION MODE The operation mode will be described in detail under the operation mode section In this mode you will perform all inspection actions such as scanning probin...

Страница 5: ...istol housing and plug in The TrisonicTM Scanning Module has a phased array of three piezoelectric transducers to pick up the airborne ultrasound This phased array focuses the ultrasound on one hot sp...

Страница 6: ...Panel will show intensity levels as dB and as a bar graph Frequency Battery Level and Display Mode The Display Mode and the Battery level indications alternate Flash Intensity levels are shown simult...

Страница 7: ...s designed to plug into a standard electrical receptacle I O PORT This is the USB port for upload download information transfer Align the pins from the cable and plug in NOTE Before downloading data b...

Страница 8: ...tenuation WTG 1 WARBLE TONE GENERATOR The WTG 1 Tone Generator is an ultrasonic transmitter designed to flood an area with ultrasound This Tone Generator is a WARBLE TONE GENERATOR This internationall...

Страница 9: ...ry charger for all Warble Tone Generators that are used with the UP9000 The line input is 120 240 VAC 60Hz 50Hz and the charging time is about 8 hours The charger comes with adaptors for multiple worl...

Страница 10: ...ne which indicates the peak intensity This is a Peak Hold function When in operation the bar graph will move up and down the scale as an indication of the amplitude of a sensed ultrasound The Peak Hol...

Страница 11: ...sensitivity is too high a blinking arrow pointing to the left will appear and there will be no numeric decibel visible on the display panel Reduce the sensitivity until the arrow disappears and the n...

Страница 12: ...r storage The lower right of the display indicates the current frequency selected for storage The lower right corner of the display indicates the Operation Mode R S or P RO SO or PO TO STORE THE READI...

Страница 13: ...blink Continue until through or until all 8 fields are filled If an error in recording a letter or number occurs click the Sensitivity Control Dial and the cursor will move to the right Continue clic...

Страница 14: ...ck the Sensitivity Control Dial and all the data will be transferred to the PC NOTE For software management refer to UltratrendTm Instructions MENU 02 SET TIME AND DATE Make sure the Ultraprobe is off...

Страница 15: ...nimum reference level set by the user This value may be any dB level above the natural 0 dB of the instrument Once set the preset level must be subtracted from the reading to determine an accurate dB...

Страница 16: ...en squeeze and hold the trigger When in the first Menu Selection Data Transfer Menu 01 the user may move to any of the other Menu Selections by spinning the Sensitivity Control up or down clockwise or...

Страница 17: ...lays and holds the peak value for comparison It changes only when a higher ultrasound level is sensed The bar graph will move up and down to display sound intensities but the Peak Hold dB reading in t...

Страница 18: ...entry is not needed select OFF TO SELECT TEXT EDITOR Make sure the Ultraprobe is off Press click both the Yellow Store button and the Sensitivity dial at the same time then squeeze and hold the trigge...

Страница 19: ...eeze and hold the trigger When in the first Menu Selection Data Transfer Menu 01 the user may move to any of the other Menu Selections by spinning the Sensitivity Control up or down clockwise or count...

Страница 20: ...SELECT FACTORY DEFAULTS Make sure the Ultraprobe is off Press click both the yellow Store button and the Sensitivity dial at the same time then squeeze and hold the trigger When in the first Menu Sele...

Страница 21: ...RUBBER FOCUSING PROBE The Rubber Focusing probe fills two functions it deflects stray ultrasounds and enhances the reception of weak airborne signals To use simply slip it over the front of the scann...

Страница 22: ...front end Align the plug located at the rear of the module with the receptacle in the front end of the Pistol Housing and plug in Start to scan the test area RAS MT The magnetically mounted transduce...

Страница 23: ...condition of the Warble Tone Generator battery set to the LOW INTENSITY position and listen to the sound through the Ultraprobe at 40 kHz A continuous warbling sound should be heard If a beeping is he...

Страница 24: ...lever Move this change lever to the middle of the lock in a way that allows it to hook behind the change notch drawing 1 Set your personal combination turning the dials to the desired combination i e...

Страница 25: ...24 Need further support Want information regarding products or training CONTACT UE Systems Inc 14 Hayes Street Elmsford NY 10523 USA T 914 592 1220 E info uesystems com W www uesystems com...

Отзывы: