10 

 

 

Once in the Sensitivity mode, turn the Sensitivity dial clockwise to increase the sensitivity 
and counter clockwise to decrease the sensitivity. 

 

The Sensitivity control dial increases/decreases the sensitivity of the instrument 
simultaneously with the sound level in the headphones. 

NOTE:

 

The instrument needs to be in range for accurate testing. 

 

If the sensitivity is too low

, a blinking arrow pointing to the right will appear and there 

will be no numeric decibel visible in the display panel.  If this occurs, increase the sensitivity 
until the arrow disappears (in low level sound environments the arrow will blink 
continuously, and it will not be possible to achieve a dB value until a higher intensity level is 
sensed). 

 

If the sensitivity is too high

, a blinking arrow pointing to the left will appear and there 

will be no numeric decibel visible on the display panel.   Reduce the sensitivity until the 
arrow disappears and the numeric decibel value is shown. 

NOTE: 

The blinking arrow indicates which direction the Sensitivity Control Dial needs to be 

turned.

 

 

The Sensitivity Control Dial controls the bar graph display. 

 

Each click of the sensitivity dial changes the volume in the headset by 1dB  

NOTE:  

The actual incident sound wave dB does not change when the sensitivity is 

adjusted. Only the headset volume. 

TO ADJUST THE FREQUENCY  

 

Look at the meter. The kHz indicator must blink to be able to tune the frequency. 

 

If it is not blinking, “Click” the Sensitivity control dial one time and the kHz indicator in the 

display panel will blink. 

 

When the kHz indicator blinks, change the frequency by rotating the Sensitivity dial up 
(clockwise) or down (counter clockwise). 

 

 

Содержание Ultraprobe 9000

Страница 1: ...0 UE SYSTEMS INC THE ULTRAPROBE 9000 MANUAL...

Страница 2: ...f Switch I O Port Battery Compartment Battery Wrist Strap Sensitivity Control Dial Storage Entry Button Head Set Jack Recharge Jack Standard Accessories Headset WTG 1 Warble Tone Generator Rubber Focu...

Страница 3: ...ext Editor Data Transfer Set Time Date dB Scale Select dB Offset Val Value This is a Reference dB Level Display Mode Calibration Due Date Text Editor Date Format Factory Defaults Exit to Program Triso...

Страница 4: ...ARE IMPORTANT TO UNDERSTAND OPERATION MODE The operation mode will be described in detail under the operation mode section In this mode you will perform all inspection actions such as scanning probin...

Страница 5: ...istol housing and plug in The TrisonicTM Scanning Module has a phased array of three piezoelectric transducers to pick up the airborne ultrasound This phased array focuses the ultrasound on one hot sp...

Страница 6: ...Panel will show intensity levels as dB and as a bar graph Frequency Battery Level and Display Mode The Display Mode and the Battery level indications alternate Flash Intensity levels are shown simult...

Страница 7: ...s designed to plug into a standard electrical receptacle I O PORT This is the USB port for upload download information transfer Align the pins from the cable and plug in NOTE Before downloading data b...

Страница 8: ...tenuation WTG 1 WARBLE TONE GENERATOR The WTG 1 Tone Generator is an ultrasonic transmitter designed to flood an area with ultrasound This Tone Generator is a WARBLE TONE GENERATOR This internationall...

Страница 9: ...ry charger for all Warble Tone Generators that are used with the UP9000 The line input is 120 240 VAC 60Hz 50Hz and the charging time is about 8 hours The charger comes with adaptors for multiple worl...

Страница 10: ...ne which indicates the peak intensity This is a Peak Hold function When in operation the bar graph will move up and down the scale as an indication of the amplitude of a sensed ultrasound The Peak Hol...

Страница 11: ...sensitivity is too high a blinking arrow pointing to the left will appear and there will be no numeric decibel visible on the display panel Reduce the sensitivity until the arrow disappears and the n...

Страница 12: ...r storage The lower right of the display indicates the current frequency selected for storage The lower right corner of the display indicates the Operation Mode R S or P RO SO or PO TO STORE THE READI...

Страница 13: ...blink Continue until through or until all 8 fields are filled If an error in recording a letter or number occurs click the Sensitivity Control Dial and the cursor will move to the right Continue clic...

Страница 14: ...ck the Sensitivity Control Dial and all the data will be transferred to the PC NOTE For software management refer to UltratrendTm Instructions MENU 02 SET TIME AND DATE Make sure the Ultraprobe is off...

Страница 15: ...nimum reference level set by the user This value may be any dB level above the natural 0 dB of the instrument Once set the preset level must be subtracted from the reading to determine an accurate dB...

Страница 16: ...en squeeze and hold the trigger When in the first Menu Selection Data Transfer Menu 01 the user may move to any of the other Menu Selections by spinning the Sensitivity Control up or down clockwise or...

Страница 17: ...lays and holds the peak value for comparison It changes only when a higher ultrasound level is sensed The bar graph will move up and down to display sound intensities but the Peak Hold dB reading in t...

Страница 18: ...entry is not needed select OFF TO SELECT TEXT EDITOR Make sure the Ultraprobe is off Press click both the Yellow Store button and the Sensitivity dial at the same time then squeeze and hold the trigge...

Страница 19: ...eeze and hold the trigger When in the first Menu Selection Data Transfer Menu 01 the user may move to any of the other Menu Selections by spinning the Sensitivity Control up or down clockwise or count...

Страница 20: ...SELECT FACTORY DEFAULTS Make sure the Ultraprobe is off Press click both the yellow Store button and the Sensitivity dial at the same time then squeeze and hold the trigger When in the first Menu Sele...

Страница 21: ...RUBBER FOCUSING PROBE The Rubber Focusing probe fills two functions it deflects stray ultrasounds and enhances the reception of weak airborne signals To use simply slip it over the front of the scann...

Страница 22: ...front end Align the plug located at the rear of the module with the receptacle in the front end of the Pistol Housing and plug in Start to scan the test area RAS MT The magnetically mounted transduce...

Страница 23: ...condition of the Warble Tone Generator battery set to the LOW INTENSITY position and listen to the sound through the Ultraprobe at 40 kHz A continuous warbling sound should be heard If a beeping is he...

Страница 24: ...lever Move this change lever to the middle of the lock in a way that allows it to hook behind the change notch drawing 1 Set your personal combination turning the dials to the desired combination i e...

Страница 25: ...24 Need further support Want information regarding products or training CONTACT UE Systems Inc 14 Hayes Street Elmsford NY 10523 USA T 914 592 1220 E info uesystems com W www uesystems com...

Отзывы: