All dimensions in inches
SPECIFICATIONS
GAIN at sum output
Input Current Range
(V/A)
Frequency Bandwidth
1 mA to 3 mA
3K
Standard
DC to 5 kHz
300
µ
A to 1 mA
10K
Optional
DC to 30 kHz
100
µ
A to 300
µ
A
30K
(Model
301DIV-30kHz)
30
µ
A to 100
µ
A
100K
10
µ
A to 30
µ
A 300K
DC drift
10
mV/C
°
for 10 V full scale
3
µ
A to 10
µ
A
1000K
Noise, rms
2 mV for 10 V full scale
Range Selector
6-position
rotary
switch
Output Impedance
Detector bias
Selectable 0 VDC (photovoltaic)
Standard
<1 ohm
or 15 VDC (photoconductive)
30 kHz version
100 ohm
Detector polarity
Selectable common anode (LSC and
Range Indicators
HI/LO signal level LEDs aid in
SC series) or common cathode
setting range switch to proper
(SPOT series)
value
Output voltage
Output
Adjustments
Sum
GAIN x (ID + IE)
CAL
Adjustable scale factor for
Position
(ID - IE)
X 10 V
position output
(ID + IE)
ZERO
Provides ± 2.5 V zero offset
Output Voltage Range
Sum
0 to 10 V
Power Requirements
Position
±10 V
Standard Model
105 to 125 V, 50/60 Hz, 5 watts
(301DIV)
Output Connectors
BNC
Optional Model
210 to 250 V, 50/60 Hz, 5 watts
Operating
0 to 50
°
C
(301DIV-230 V)
Temperature
UDT Instruments • 8581 Aero Drive • San Diego, CA 92123
Phone (858) 279-8035 Fax (858) 576-9286