background image

PAGE 28 OF 33 

1219 IH-8105

COMMAND FUNCTION

LEVEL 0

@

Reset The scale.

I0

Inquiry of all available SICS commands.

I1

Inquiry of SICS level and SICS versions.

I2

Inquiry of scale data.

I3

Inquiry of scale software version.

I4

Inquiry of serial number.

S

Send stable weight value.

SI

Send weight value immediately.

SIR

Send weight value repeatedly.

Z

Zero the scale.

ZI

Zero immediately.

LEVEL 1

D

Write text into display.

DW

Weight display.

SR

Send and repeat stable weight value.

T

Tare.

TA

Tare value.

TAC

Clear tare.

TI

Tare immediately.

LEVEL 2

C2

Calibrate with the external calibration weight.

C3

Calibrate with the internal calibration weight.

I10

Inquire or set scale ID.

I11

Inquire of scale type.

P100

Print out on the printer. 

P101

Print out stable weight value.

P102

Print out current weight value immediately.

SIRU

Send weight value in the current unit immediately and repeat.

SIU

Send weight value in the current unit immediately.

SNR

Send stable weight value and repeat after every weight change.

SNRU

Send stable weight value in the current unit and repeat after every weight change.

SRU

Send weight value in the current unit and repeat.

ST

After pressing the transfer key, send the stable weight value.

SU

Send stable weight value in the current unit.

LEVEL 3

M01

Weighing mode.

M02

Stability setting.

M03

Autozero function.

M19

Send calibration weight.

M21

Inquire/set weight unit.

PRN

Print out at every printer interface.

RST

Restart.

SFIR

Send weight value immediately and repeat quickly.

SIH

Send weight value immediately in high resolution.

SWU

Switch weight unit.

SX

Send stable data record.

SXI

Send data record immediately.

SXIR

Send data record immediately and repeat.

U

Switch weight unit.

MT-SICS Commands

SERIAL COMMUNICATION AND PRINTER SETUP CONTINUED

Содержание OHAUS DEFENDER 5000

Страница 1: ...its Print Short press Send the current display value to the RS 232 port or option when properly enabled Long Press Change the current weighing unit Press and hold the key to scroll through the list of...

Страница 2: ...1 a Switch between three input modes numeric lowercase and uppercase input OVERVIEW OF CONTROLS SETUP 1 Remove shipping spacers and place the weighing pan securely over the rubber load pads on the top...

Страница 3: ...ate Edit Template Edit String Reset WiFi Bluetooth WiFi MAC Address Network Port DHCP IP Address Gateway DNS Subnet Mask Alternate Command Reset Bluetooth Device Name Print Setup Assignment Print Opti...

Страница 4: ...FT set to ON 4 Allow the scale to warm up for approximately five minutes after stabilizing to room temperature 5 To abort calibration press the softkey corresponding to the X icon anytime during the c...

Страница 5: ...to enter the Calibration submenu Scroll to Linearity Calibration using the softkey corresponding to the icon See Figure 14 3 Press the softkey corresponding to the icon to initiate Linearity Calibrati...

Страница 6: ...10 19 02 7 7 6 6 5 5 4 4 3 3 2 19 02 20 45 8 7 7 6 6 5 5 4 4 3 3 20 45 22 22 8 8 7 7 6 6 5 5 4 4 3 22 22 23 54 9 8 8 7 7 6 6 5 5 4 4 23 54 25 21 9 9 8 8 7 7 6 6 5 5 4 25 21 26 45 10 9 9 8 8 7 7 6 6 5...

Страница 7: ...6 26 25 85 45 90 00 31 30 30 29 29 28 28 27 27 26 26 OPERATION CONTINUED TABLE OF GEO VALUES CONTINUED SETUP MENU When the Indicator connects to a scale base for the first time enter this menu to set...

Страница 8: ...tion for the second weighing range of the scale If Single interval is enabled this parameter will not be shown NOTE Graduation settings are limited to values from Capacity divided by 600 to Capacity d...

Страница 9: ...intain zero until a change of one division per second is exceeded 3d Display will maintain zero until a change of three divisions per second is exceeded AUTO DIM Set the display backlight functionalit...

Страница 10: ...late Discrete Output 1 Off Overload Underload Zero Discrete Output 2 Off Overload Underload Zero Discrete Output 3 Off Overload Underload Zero Discrete Output 4 Off Overload Underload Zero Counting En...

Страница 11: ...actor multiplier 3 divide the Factor by 1 000 1x10 3 2 divide the Factor by 100 1x10 2 1 divide the Factor by 10 1x10 1 0 multiply the Factor by 1 1x100 1 multiply the Factor by 10 1x101 2 multiply th...

Страница 12: ...ss the button Mode to select an activated application The current application will be shown in the upper left corner of the home screen The Indicator incorporates the following applications WEIGHING U...

Страница 13: ...esults 1 To view accumulation results press the softkey corresponding to the icon The Accumulate Result screen is displayed See Figure 31 2 To clear accumulation results press the Off On CLR button Se...

Страница 14: ...umber of pieces SET THE AVERAGE PIECE WEIGHT APW NOTE It is recommended that the APW is larger than 1d If APW is between 0 05d and 1d a warning message will be displayed and the information line will...

Страница 15: ...late Off Automatic Manual Off The icon does not display Manual The softkey displays User can press the key to accumulate the stable weight Automatic The icon displays The weight will be accumulated au...

Страница 16: ...Figure 54 CHECK Check is used to compare the weight or pieces of a sample against target limits Press the Mode Library button until the icon corresponding to Check is displayed on the screen Two diffe...

Страница 17: ...and Over Accept Under Over Off No beep Under and Over Accept Under Over Beep when reaching the selected check point CHECK COUNTING 1 Press the button and select Check Mode to Check Counting See Figure...

Страница 18: ...e stable weight Automatic The icon displays The weight will be accumulated automatically NOTE The load to be accumulated has to be 10d The next accumulation can only start once the pan is cleared When...

Страница 19: ...e Figure 65 2 Press the softkey corresponding to the icon for confirmation See Figure 66 Method 2 1 Press the softkey corresponding to the icon to enter the submenu for setting reference weight See Fi...

Страница 20: ...list and highlight Application Mode See Figure 72 2 Enter this submenu by pressing the button corresponding to the icon See Figure 72 3 In the Application Mode menu enter the Percent submenu See Figu...

Страница 21: ...e placed on the pan first After the terminal has done taring automatically place load on the pan to start dynamic weighing Start Weight 3 000 with current unit If the operation type is Semi Automatic...

Страница 22: ...s setup will only work when the I O Option board has been installed The option I O board provides two isolated inputs and four dry contact normally open relay outputs which can be used for simple pro...

Страница 23: ...table Only Off On LFT Force On Auto On Stable Mode Load Load and Zero Auto On Accept Interval Time 1 50 000 MT Continuous OH Continuous SICS Reference Balance Select Template Simple Custom 1 Custom 2...

Страница 24: ...lected and weighing mode is Check values will be printed when weight is accepted ACCEPT Printing occurs each time the display is within the Checkweigh accept range and stability criteria are met INTER...

Страница 25: ...minate a template an End of Template field must be included All fields after the End of Template field will be ignored Figure 87 Figure 88 ITEM LENGTH ITEM LENGTH 3 spaces 3 String 1 Not fixed up to 4...

Страница 26: ...structure 3 Displayed weight Either gross or net weight Six digits no decimal point or sign Insignificant leading zeroes are replaced with spaces 4 Tare weight Six digits of tare weight data No decim...

Страница 27: ...re xxx is the value under current unit CO xxx Set Over Limit only in Check mode where xxx is the value under current unit x Set Counting APW x in grams only in Counting or Check Counting mode must hav...

Страница 28: ...nt weight value immediately SIRU Send weight value in the current unit immediately and repeat SIU Send weight value in the current unit immediately SNR Send stable weight value and repeat after every...

Страница 29: ...ify communication by pressing the print button NOTE If HyperTerminal is set up properly the value on the display will be shown in the window CONNECTING TO A SERIAL PRINTER Connect the cable supplied w...

Страница 30: ...used to separate the other fields The Weight1 field is four right justified characters If the value is negative the character is located at the immediate left of the most significant digit The Unit1...

Страница 31: ...k on the terminal BATTERY OPERATION During operation battery icon indicates charge level remaining The battery icon is located above weighing unit on indicator display See Figure 94 NOTE When battery...

Страница 32: ...verification procedure SEALING The local weights and measures official or authorized service agent must apply a security seal to prevent tampering with the settings Refer to the illustrations below fo...

Страница 33: ...ected menu to off in the Lock menu Lockout Switch on the circuit board may need to be set to the off position Error 8 1 Weight reading exceeds Power On Zero limit Remove load from scale Recalibrate sc...

Отзывы: