Doc
No.
FM0706
issue
D
Page
2
INITIAL
SET
‐
UP
From
new
you
will
need
to
open
the
battery
compartment
and
remove
the
battery
isolation
strip.
The
display
will
show
“
BOOT
”
followed
by
the
software
version
e.g.
“
SW
‐
007
”
then
the
battery
charge
level
“
B
100%”
(full
charge).
When
the
display
shows
“
CLOCK
”
and
flashes
“
00:00
”
the
unit
is
ready
to
program.
There
are
2
programming
menus;
Menu
A
contains
the
basic
settings
and
Menu
B
contains
more
advanced
functions.
See
below;
MENU
A
SETTINGS
Press
button
1
to
cycle
through
the
following
functions;
CLOCK
DATE
DOSES
ALARMS
Press
and
hold
button
1
for
3
seconds
to
exit
the
menu
and
revert
to
normal
operation.
Setting
the
Clock
Press
button
1
once
and
CLOCK
is
displayed.
Press
button
2/3
to
set
the
time
(00:00
to
23:59)
Note:
press
and
hold
for
fast
forward/reverse
Press
button
1
to
end
and
DATE
will
be
displayed.
Setting
the
Date
The
DATE
display
will
change
to
YY
after
a
few
seconds.
Press
button
2/3
to
set
the
YEAR
(00
to
99)
Press
button
1
and
MM
will
be
displayed
Press
button
2/3
to
set
the
MONTH
(01
to
12)
Press
button
1
and
DD
will
be
displayed
Press
button
2/3
to
set
the
DATE
(01
to
31)
Press
button
1
to
end
and
DOSES
will
be
displayed.
Setting
the
Number
of
Doses
per
Day
The
DOSES
display
will
change
to
display
the
number
of
doses
per
day
after
a
few
seconds.
Press
button
2/3
to
set
the
doses
per
day
(00
to
24)
Press
button
1
to
end
and
ALARMS
will
be
displayed.
Setting
the
Alarm
Times
for
each
Dose
The
ALARMS
display
will
change
to
display
the
time
for
the
first
dose
after
a
few
seconds.
Press
button
2/3
to
set
the
alarm
time
for
the
first
dose
Press
button
1
to
end
then
set
alarm
time
2
Repeat
for
each
alarm
time,
the
number
of
alarm
times
is
determined
by
the
“number
of
doses
per
day”
set
above.
MENU
B
SETTINGS
Press
and
hold
button
3
until
BEFORE
is
displayed
then
press
button
1
to
cycle
through;
AFTER
SOUND
LIGHT
FILLED
BATT
SECURE
HOURS
DLS
START
STOP
RECORD
Press
and
hold
button
1
for
3
seconds
to
exit
the
menu
and
revert
to
normal
operation.
Setting
the
Before
Time
The
BEFORE
time
allows
medication
to
be
taken
up
to
5
hours
earlier
than
the
next
alarm
time.
Press
and
hold
button
3
until
BEFORE
is
displayed
Press
button
2/3
to
set
the
before
time
(00:00
=
not
required,
to
05:00
max)
Press
button
1
to
end
and
AFTER
will
be
displayed
If
this
option
is
set
the
user
may
access
their
medication
in
advance
by
turning
the
pill
dispenser
upside
down
then
back
upright
again
twice
in
succession,
the
tray
will
rotate
one
step
and
the
next
dose
will
become
available.
Setting
the
After
Time
The
AFTER
time
sets
how
long
the
alarm
sounds
and
the
pill
shutter
remains
unlocked
after
the
medication
is
due.
After
AFTER
is
displayed;
Press
button
2/3
to
set
the
after
time
(00:05
to
05:00)
Press
button
1
to
end
and
SOUND
will
be
displayed.
If
half
the
AFTER
time
expires
an
alarm
call
will
be
transmitted.
If
the
full
AFTER
time
expires
another
alarm
call
will
be
transmitted
and
the
pill
shutter
will
lock
shut.
Setting
the
Alarm
Sound
The
alarm
SOUND
will
play
to
alert
the
user
when
it
is
time
to
take
their
medication.
It
will
sound
every
5
seconds
for
the
first
minute,
then
every
8
seconds
up
to
1
hour
then
once
a
minute
until
the
medication
is
dispensed
or
the
AFTER
time
is
reached,
maximum
5
hours.
After
SOUND
is
displayed;
Press
buttons
2/3
to
select
the
preferred
sound
(01
to
04)
Select
00
for
no
sound
Press
button
1
to
end
and
LIGHT
will
be
displayed.
Setting
the
Light
On/Off
The
LIGHT
will
flash
to
alert
the
user
when
it
is
time
to
take
their
medication.
It
will
flash
every
5
seconds
for
the
first
minute,
then
every
8
seconds
up
to
1
hour
then
once
every
16
seconds
until
the
medication
is
dispensed
or
the
AFTER
time
is
reached,
maximum
5
hours.
After
LIGHT
is
displayed;
Press
button
2
for
no
light
(display
shows
OFF)
Press
button
3
for
light
(display
shows
ON)
Press
button
1
to
end
and
FILLED
will
be
displayed.
Setting
the
Filled
Quantity
This
is
the
number
of
doses
placed
in
the
medication
tray
when
filled
and
allows
the
DOSES
LEFT
display
to
be
shown.
After
FILLED
is
displayed;
Press
button
2/3
to
set
the
FILLED
quantity
(00
to
28)
Press
button
1
to
end
and
BATT
will
be
displayed.
Setting
the
Battery
Type
It
is
possible
to
use
4
x
AA
size
alkaline
or
4
x
AA
size
rechargeable
batteries
in
the
‘Advance’
Pill
Dispenser.
After
BATT
is
displayed;
Press
button
2
to
select
alkaline
(
ALC
)
Press
button
3
to
select
rechargeable
(
NIMH
)
Press
button
1
to
end
and
SECURE
will
be
displayed.
Setting
the
Secure
Mode
With
SECURE
mode
ON
the
pill
shutter
will
be
locked
except
at
alarm
time
and
for
the
subsequent
after
time.
With
SECURE
mode
OFF
the
shutter
does
not
lock.
After
SECURE
is
displayed;
Press
button
2
to
select
secure
mode
ON
Press
button
3
to
select
secure
mode
OFF
Press
button
1
to
end
and
HOURS
will
be
displayed.
Setting
the
Hours
Mode
The
clock
can
be
set
to
display
the
time
in
12
hour
AM/PM
mode
or
24
hour
mode.
After
HOURS
is
displayed;
Press
button
2
to
select
12
hour
clock
(AM/PM)
Press
button
3
to
select
24
hour
clock
Press
button
1
to
end
and
DLS
will
be
displayed.
Setting
the
Daylight
Saving
Mode
To
clock
can
be
set
to
automatically
adjust
+/
‐
1
hour
every
March/October
for
GMT/BST.
After
DLS
is
displayed;
Press
button
2/3
to
select
OFF/EU/US
Set
EU
for
Europe,
US
for
America
or
OFF
to
disable.
Press
button
1
to
end
and
START
will
be
displayed.