Class B: General description
Class B (that conform to European standard) mainly provides deeper vacuum
in the autoclave chamber than Class S, and therefore include more complete
air removal from the most difficult load types. It allows the system to perform
vacuum in difficult hollow or porous instruments and perform effective
sterilization of these materials.
Class S: General description
Class S used in the USA obtain weaker vacuum level and is less effective
than Class B in the sterilization of hollow instruments.
The following tables provide:
Table A
for types of sterilization cycles
Table B
provides the T-EDGE 10, Class B sterilization programs.
Table C
provides the T-EDGE 10, Class S sterilization programs.
Table A — Types of sterilization cycles
Class B
Class S
The sterilization of products as
represented by the test loads in the
standard.
For products that lie within the limits
specified for the relevant test loads,
this includes:
solid products
porous products and lumen devices
wrapped (single and multiple-layer)
or non-wrapped
The sterilization of products as
specified by the manufacturer of the
sterilizer including non-wrapped
solid products and at least one of
the following:
porous products, small porous
items
lumen devices, bowls and
receivers
single-layer wrapped products
multiple-layer wrapped products
Page 36
Содержание AMS10-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Страница 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Страница 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Страница 85: ...Page 78...
Страница 104: ...Cat No MAN205 0502015EN Rev H COPYRIGHT...