1.4
General Description of the Device
The autoclave is fully automatic (a computerized control unit ensures a fully
automatic sterilization cycle, control and monitoring of physical parameters
and a clear documentation of the sterilization cycle. Drying is performed with
the door closed)
This autoclave uses steam as a sterilizing agent.
The steam is produced by warming up a controlled amount of water inserted to
a pipe heating element, and then to the chamber. This technique saves energy
and water consumption.
The autoclave is equipped with a Pipe heating element and with chamber
heaters to maintain the steam inside the chamber.
The autoclave is equipped with a vacuum system, which supports and
improves:
l
Removal of residual air from packs and porous load and most
kinds of tubes (rubber, plastic etc.) by vacuum at the first stage of
the cycle.
l
Steam penetration into the load; resulting in effective sterilization.
l
Temperature uniformity.
l
Post sterilization drying phase
A touchscreen is used for monitoring and control purposes.
The device has a built -in USB port to enable the operation of an external
optional barcode printer:
l
The barcode printer can print labels with a unique cycle ID
barcode, operator name, sterilization and expiry dates
l
One barcode printer can be connected to the machine.
l
The printer connection to the machine, by using a USB socket,
with a dedicated cable.
l
Barcode printer power supply voltage can range between 100-
240V.
l
A barcode printer is an optional addition to the autoclave
The device features built-in memory to record up to 999 sterilization cycles.
These can be exported to a USB device to be transferred to a PC.
The device has a built-in Network Port for use with optional Tuttnauer’s
R.PC.R software when connected to your local network.
The autoclave has two optional configurations (available upon request):
Page 5
Содержание AMS10-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Страница 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Страница 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Страница 85: ...Page 78...
Страница 104: ...Cat No MAN205 0502015EN Rev H COPYRIGHT...