1.10.2
T-Edge10 fit into sterilization center cabinet - recommended
dimensions:
To allow adequate cooling & ventilation of the autoclave placed inside the
sterilization center Cabinet, the following T-Edge10 Cabinet sizes are
recommended:
Width - 22.85”- 23.65”/58-60cm; Height – 22.25”- 22.65”/56.5-57.5cm; Min.
Depth - 24.9”-25.2”/63.2-64cm
Note:
The imperial dimensions in inches/lbs are approximated to less than 3
percent.
1.10.3
Device Properties
Table 2 - Properties of the device
Property
Dimension
External
size
Width
~19.7"(50cm)
Height
~19"(49 cm)
Depth
~22.8" (58 cm) supporting common install base carry
a ~23" (60 cm) counter top
Chamber
Diameter
~10"(25 cm)
Depth
~18"(46 cm)
Volume
~778 Ounces (23lit)
Usable chamber
space
75% (~575 Ounces / ~17 L)
Max. Allowable Working
Pressure (MAWP)
~40.6PSI (2.8 bar)
Maximum load per item
~0.45 lbs. (0.2kg)
Maximum
load per tray
Unwrapped
~2.67 lbs. ( 1.2kg)
Wrapped
~1.6 lbs (0.72kg)
Maximum
Solid load
Unwrapped
~13 lbs.(6kg)
Wrapped
~7.7 lbs.(3.5kg)
Maximum textile load
~3.3 lbs.(1.5kg)