5.2.1 User Manual Isoforce
REV 5
01st July 2014
page 63 von 77
TUR Therapietechnik GmbH | Grubenstr. 20 | 18055 Rostock | Germany
Attachments Needed
Dynamometer: ADL02, ADW1
Chair: ADL01, ADS4
Put the mechanical stops to “B” blue and “B” red.
Collect angle/elbow attachment and the handle grip and combine them as shown in the picture
above
Combine ADL1 with the ADS4 and put them on the chair input adapter. The forearm stabilizer
should be at the side of the wrist to be tested.
Attach the adapter to the dynamometer by sliding it in the input arm from the top.
Ask the subject to lay on the positioning chair or seat on the chair.
Ask the subject to hold the grip.
Set the axis of the joint rotation.
Check the axis of rotation by asking your subject to perform full wrist ulnar and radial deviation.