5.2.1 User Manual Isoforce
REV 5
01st July 2014
page 52 von 77
TUR Therapietechnik GmbH | Grubenstr. 20 | 18055 Rostock | Germany
Attachments Needed
Dynamometer: Angle Adaptor (COMB05) comprised of AD008+AD012
Put the mechanical stops to “B” blue and “B” red.
Collect angle/elbow attachment and the handle grip. Combine them to form the attachment
COMB05.
Attach the COMB05 to the dynamometer by sliding it in the input arm.
Adjust the chair and the dynamometer in accordance to the quick reference above.
Ask the subject to lay on the positioning chair.
Ask the subject to hold the handle and bring his/her shoulder in 90° flexion.
Set the axis of the joint rotation. The axis of rotation for shoulder is through the acromiom, and it
should be right opposite to the coloured dot on the input arm of the dynamometer. To succeed
the alignment you might need to make alterations to the positions of the chair and the
dynamometer.
The length of the attachment might need to be adjusted.
Check the axis of rotation by asking your subject to perform a horizontal abduction and adduction
movement.
Opposite Side
Ask the subject to leave the handle.
While the subject still lays on the chair, slide the chair away from the dynamometer.
Remove the shoulder attachment from the dynamometer input arm and put it back turning it
around.
Adjust the chair and the dynamometer in accordance to the quick reference above.
Follow the steps 6 -9 as described above.
Considerations
The movement is performed more comfortable if the elbow is not fully flexed