5.2.1 User Manual Isoforce
REV 5
01st July 2014
page 61 von 77
TUR Therapietechnik GmbH | Grubenstr. 20 | 18055 Rostock | Germany
Extension, however optimal ROM for the isokinetic testing of the wrist is between 40
degrees extension and 60 degrees flexion.
Isoforce offers one positioning possibility, in seated position.
Muscles Involved: The muscles that flex the hand at the wrist are the flexor carpi radialis, flexor carpi
ulnaris, and palmaris longus and the muscles that extend the hand at the wrist are the extensor carpi
radialis longus, extensor carpi radialis brevis and extensor carpi ulnaris.
Attachments Needed
Dynamometer: ADL02, ADW1
Chair: ADL01, ADS4
Put the mechanical stops to “B” blue and “B” red.
Collect angle/elbow attachment and the handle grip and combine them as shown in the picture
above
Combine ADL1 with the ADS4 and put them on the chair input adapter. The forearm stabilizer
should be at the side of the wrist to be tested.
Attach the adapter to the dynamometer by sliding it in the input arm from the top.
Ask the subject to lay on the positioning chair or seat on the chair.