Operating Facilities 

Noise Filter (Input A only) 

The Filter key controls a low pass filter, with a cut-off frequency of about 50kHz, to assist in 
obtaining stable readings at low frequencies. 

Hold 

Pressing the Hold key will hold the current measured value in the display, with the Hold indicator 
on, until the Hold key is pressed again. The measurement continues in the background when 
Hold is on. A long press on the Hold key clears old data and restarts the measurement. 

Intelligent Power Switching 

The unit automatically selects the best available power source of AC adaptor, USB or battery. 
Intelligent switching avoids discharging the battery overnight when operated from externally 
switched AC power. 

A press-to-measure facility allows a quick measurement to be made by pressing a function select 
key which will power the instrument up in the corresponding function. The instrument will 
automatically switch off 15 seconds after the last key-press. 

Remote Control 

All capabilities can be controlled remotely and measurements read through a USB port.  

The instrument can be powered (but the battery cannot be charged) by the USB host. 

Interface: 

Serial port emulation over USB. 

Current consumption: 

< 100mA (<5mA if AC adaptor power is present) 

Command set: 

Instrument specific. TF830 and TF930 compatible. 

Power Requirements 

The instrument has fixed internal rechargeable batteries and is supplied with a universal voltage 
external mains adaptor with interchangeable UK, Euro, Australian and US power connectors. 

Battery Type: 

Three 2500mAh NiMH cells. 

Battery Operating Life:  Typically 24 hours 

Low Battery Indicator: 

‘Lo Bat' shows in display when approximately 10% of battery life remains. 

Recharge Time: 

< 4 hours 

Adaptor Supply range:  85 to 240V, 50 or 60 Hz, 

Power consumption: 

5W max at DC input to unit; 15VA max at AC adaptor input (charging). 

General 

Display: 

10 digit LCD, 12.5mm high (0.5”).  Annunciators show input 
configuration, operating mode, measurement units and gate time. 

Operating Range: 

+5°C to +40°C, 20% to 80% RH 

Storage Range: 

20°C to +60°C 

Environmental: 

Indoor use at altitudes up to 2000m, Pollution Degree 2 

Size: 

260mm(W) x 88mm(H) x 235mm(D) 

Weight: 

1050 gms (plus 170 gms AC adaptor) 

Electrical Safety: 

Complies with EN61010-1 

EMC: 

Complies with EN61326 

 

Содержание TF930

Страница 1: ...TF930 TF960 3GHz 6GHz Universal Counters SERVICE GUIDE...

Страница 2: ...1 Table of Contents Specifications 2 Safety 5 General 6 Frequency Calibration 7 C Input Threshold Adjustment 8 Diagnostic Notes 9 Battery replacement 11 Parts List 12...

Страница 3: ...MHz at optimum threshold adjustment Input B Input Impedance 50 nominal AC coupled Frequency Range 80MHz to 3GHz Sensitivity 12mVrms 80MHz to 2GHz 25mVrms to 2 5GHz 50mVrms to 3GHz Maximum Input Signal...

Страница 4: ...aging Automatic within measurement time selected up to 50 pulses Resolution 20ns for one pulse up to 1ns or 10 digits with multiple pulse averaging 0 01 for Ratio H L and Duty Cycle Total Count Input...

Страница 5: ...asurements read through a USB port The instrument can be powered but the battery cannot be charged by the USB host Interface Serial port emulation over USB Current consumption 100mA 5mA if AC adaptor...

Страница 6: ...safety under all other circumstances it is essential that no input is connected to a voltage above 30Vdc or 30Vrms with respect to earth ground which is the limit of Safe Extra Low Voltage SELV by IE...

Страница 7: ...Before removing the rear panel or the Battery USB PCB note the position of the grounding spring attached to the USB socket The PCB is attached to the case lower with 4 screws To avoid short circuiting...

Страница 8: ...andard source subject to any pre calibration inaccuracy If the reference signal is large it may be necessary to select the 5 1 attenuator The calibration setting is adjusted by pressing the Frequency...

Страница 9: ...nerator with reasonably accurate calibrated level around 5 to 15mV A microwave grade N type to N type coaxial lead Do not use a BNC type with adapters Procedure To adjust the setting first dismantle t...

Страница 10: ...resort protection against fire and any failure indicates that another major fault has occurred Note in particular that this component MUST NOT be hand soldered as that will almost certainly cause an...

Страница 11: ...unless the port has been properly enumerated by the host PC If the PC can detect the USB device but functional control of the counter is not possible then provided the baud rate etc are correctly con...

Страница 12: ...ctly oriented they all point in the same direction and that the tags are aligned with the correct connecting pads Secure the cells to the PCB with tie wraps only sufficiently tight to hold the cells d...

Страница 13: ...210 BEZEL BENCH CASE 2 33536 4380 CASE UPPER BENCH CASE 2 33536 4390 CASE LOWER BENCH CASE 2 33331 9790 REAR PANEL 37113 2183 KEYCAP 8X6MM GREY3 TF 37151 0483 KNOB 21MM D SHAFT GREY3 38611 0010 COLLAR...

Страница 14: ...idgeshire PE29 7DR England United Kingdom Telephone 44 0 1480 412451 Fax 44 0 1480 450409 International web site www aimtti com UK web site www aimtti co uk Email info aimtti com Aim Instruments and T...

Отзывы: