Timebase 

Measurement Clock: 

50MHz. 

Internal Reference oscillator:  

10MHz TCXO with electronic calibration adjustment. 

Oscillator Temperature Stability: 

Better than ± 1ppm over rated temperature range. 

Initial Oscillator Adjustment Error: 

< ± 0.2ppm at 21ºC.  

Oscillator Ageing Rate: 

< ± 1ppm first year. 

Calibration adjustment range: 

> ± 8ppm. 

Measurement Functions 

Frequency (Inputs A, B or C) 

A Input Frequency Range: 

< 0.001Hz (DC coupled) to >125MHz 

B input Frequency Range: 

80MHz to >3000MHz. 

C input Frequency Range: 

<2Ghz to >6GHz. 

Resolution: 

up to 10 digits (see below) or 0.001Hz 

Period (Inputs A, B or C) 

A Input Period Range:  

8ns to 1000s (DC coupled) 

B input Period Range: 

0.333ns to 12.5ns 

C input Period Range: 

0.166ns to 0.5ns 

Resolution: 

up to 10 digits (see below) 

Pulse Width Modes (Input A only) 

Functions: 

Width high or low, ratio H

:

L (high time to low time) or duty cycle. 

Pulse Width Range: 

40ns to 1000s 

Averaging: 

Automatic within measurement time selected, up to 50 pulses. 

Resolution: 

20ns for one pulse; up to 1ns or 10 digits with multiple pulse 
averaging. 0.01% for Ratio H

:

L and Duty Cycle. 

Total Count (Input A only) 

Count range: 

1 to 9 999 999 999 

Minimum pulse width: 

8ns 

Frequency Ratio B:A 

Resolution: 

Equal to the resolution of the two frequency measurements. 
If the ratio exceeds ten digits, displays six digits plus exponent. 

Measurement Time 

Selectable as 100s, 10s, 1s or 0.3s.  The instrument displays the average value of the input 
signal over the measurement time selected, updated every 2s, 1s, 0.5s or 0.3s respectively. The 
hardware captures the count values and continues measuring without any dead time. 

Resolution 

The displayed resolution depends upon measurement time and input frequency. The basic 
resolution of period is 8 digits for every 2 seconds of measurement time. Frequency resolution is 
the reciprocal of period resolution. Usable resolution can be reduced by noise at low frequencies. 

Accuracy 

Measurement accuracy is timebase ac measurement reso 2 counts. 

Содержание TF930

Страница 1: ...TF930 TF960 3GHz 6GHz Universal Counters SERVICE GUIDE...

Страница 2: ...1 Table of Contents Specifications 2 Safety 5 General 6 Frequency Calibration 7 C Input Threshold Adjustment 8 Diagnostic Notes 9 Battery replacement 11 Parts List 12...

Страница 3: ...MHz at optimum threshold adjustment Input B Input Impedance 50 nominal AC coupled Frequency Range 80MHz to 3GHz Sensitivity 12mVrms 80MHz to 2GHz 25mVrms to 2 5GHz 50mVrms to 3GHz Maximum Input Signal...

Страница 4: ...aging Automatic within measurement time selected up to 50 pulses Resolution 20ns for one pulse up to 1ns or 10 digits with multiple pulse averaging 0 01 for Ratio H L and Duty Cycle Total Count Input...

Страница 5: ...asurements read through a USB port The instrument can be powered but the battery cannot be charged by the USB host Interface Serial port emulation over USB Current consumption 100mA 5mA if AC adaptor...

Страница 6: ...safety under all other circumstances it is essential that no input is connected to a voltage above 30Vdc or 30Vrms with respect to earth ground which is the limit of Safe Extra Low Voltage SELV by IE...

Страница 7: ...Before removing the rear panel or the Battery USB PCB note the position of the grounding spring attached to the USB socket The PCB is attached to the case lower with 4 screws To avoid short circuiting...

Страница 8: ...andard source subject to any pre calibration inaccuracy If the reference signal is large it may be necessary to select the 5 1 attenuator The calibration setting is adjusted by pressing the Frequency...

Страница 9: ...nerator with reasonably accurate calibrated level around 5 to 15mV A microwave grade N type to N type coaxial lead Do not use a BNC type with adapters Procedure To adjust the setting first dismantle t...

Страница 10: ...resort protection against fire and any failure indicates that another major fault has occurred Note in particular that this component MUST NOT be hand soldered as that will almost certainly cause an...

Страница 11: ...unless the port has been properly enumerated by the host PC If the PC can detect the USB device but functional control of the counter is not possible then provided the baud rate etc are correctly con...

Страница 12: ...ctly oriented they all point in the same direction and that the tags are aligned with the correct connecting pads Secure the cells to the PCB with tie wraps only sufficiently tight to hold the cells d...

Страница 13: ...210 BEZEL BENCH CASE 2 33536 4380 CASE UPPER BENCH CASE 2 33536 4390 CASE LOWER BENCH CASE 2 33331 9790 REAR PANEL 37113 2183 KEYCAP 8X6MM GREY3 TF 37151 0483 KNOB 21MM D SHAFT GREY3 38611 0010 COLLAR...

Страница 14: ...idgeshire PE29 7DR England United Kingdom Telephone 44 0 1480 412451 Fax 44 0 1480 450409 International web site www aimtti com UK web site www aimtti co uk Email info aimtti com Aim Instruments and T...

Отзывы: