  4 

Safety 

This instrument is Safety Class III according to IEC classification and has been designed to meet 
the requirements of EN61010-1 (Safety Requirements for Electrical Equipment for Measurement, 
Control and Laboratory Use). This instrument has been tested in accordance with EN61010-1 and 
has been supplied in a safe condition. This instruction manual contains some information and 
warnings which have to be followed by the user to ensure safe operation and to retain the 
instrument in a safe condition.  This instrument has been designed for indoor use in a Pollution 
Degree 2 environment in the temperature range 5°C to 40°C, 20%-80% RH (non-condensing).  
It may occasionally be subjected to temperatures b5°C and 

10°C without degradation of 

its safety. Do not operate while condensation is present. Use of this instrument in a manner not 
specified by these instructions may impair the safety protection provided.  Do not operate the 
instrument outside its rated supply voltages or environmental range.  

WARNING! 

All accessible parts will be at the same voltage potential as the outer of the BNC input 

sockets: to maintain user safety it is therefore essential that neither input is connected to a voltage 

above 30Vdc or 30Vrms with respect to ground 

.   Note that although the inputs will withstand 

accidental connection to an AC line voltage up to 250V rms, 50/60Hz, users will be at risk if the 
instrument 'ground' is connected to such hazardous  voltages.  The apparatus shall be 
disconnected from all voltage sources before it is opened for any adjustment, replacement, 
maintenance or repair, including replacement of the battery. 

Any adjustment, maintenance and repair of the opened instrument under voltage shall be avoided 
as far as possible and, if inevitable, shall be carried out only by a skilled person who is aware of the 
hazard involved. 

Do not wet the instrument when cleaning it. The following symbol is used on the instrument and in 
this manual: 

 

meaning  CAUTION. Damage to the instrument may occur if these 
precautions are ignored. 

EMC 

This instrument has been designed to meet the requirements of the EMC Directive 2004/108/EC. 
Compliance was demonstrated by meeting the test limits of the following standards: 

Emissions 

EN61326 (2006) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use.  Test limits used were:     Radiated: Class B 

Immunity 

EN61326 (2006) EMC product standard for Electrical Equipment for Measurement, Control and 
Laboratory Use. 

Test methods, limits and performance achieved are shown below (requirement shown in brackets): 

a) 

EN61000-4-2 (2009) Electrostatic Discharge : 4kV air, 4kV contact, Performance A (B). 

b) 

EN61000-4-3 (2006) Electromagnetic Field: 
3V/m, 80% AM at 1kHz, 80MHz - 1GHz: Performance A (A) and 1.4GHz to 2GHz: 
Performance A (A); 1V/m, 2.0GHz to 2.7GHz: Performance A (A). 

c) 

EN61000-4-4 (2004) Fast Transient, 0.5kV peak (signal inputs), Performance B (B). 

According to EN61326-1 the definitions of performance criteria are: 

Performance criterion A:

  ‘During test normal performance within the specification limits.’ 

Performance criterion B:

  ‘During test, temporary degradation, or loss of function or performance 

which is self-recovering. 

Содержание PFM3000

Страница 1: ...PFM3000 Portable Frequency Meter INSTRUCTION MANUAL...

Страница 2: ...old function is available which allows any reading to be frozen and displayed indefinitely Input A is a high impedance input 1M for frequencies in the range 3Hz to 125MHz Input B is a nominal 50 input...

Страница 3: ...accuracy is the sum of timebase accuracy and measurement resolution plus one count TIMEBASE Crystal Oscillator Frequency 10MHz Initial Oscillator Adjustment Error 2ppm closed case adjustment by user...

Страница 4: ...dc 30V rms 50 60Hz with respect to ground reducing to 1Vrms above 1MHz Note that the inputs will not be damaged if subjected to an accidental short term connection to a 50 60Hz line voltage not exceed...

Страница 5: ...connected from all voltage sources before it is opened for any adjustment replacement maintenance or repair including replacement of the battery Any adjustment maintenance and repair of the opened ins...

Страница 6: ...by conformance to the following specifications which have been listed in the Official Journal of the European Communities EMC Emissions EN61326 1 2006 Radiated Class B Immunity EN61326 1 2006 Immunit...

Страница 7: ...ting or restarting the measurement gives a true result with at least 6 digits after 0 3s followed by at least 7 digits after 1s and finally at least 8 digits after 10s if 10s measurement is selected T...

Страница 8: ...correct reading Signals with slow edge speeds such as low frequency sinewaves will show some jitter on the reading because of random noise TIMEBASE OTHER ACCURACY CONSIDERATIONS The instrument has an...

Страница 9: ...limentation nominales ou de la gamme des conditions ambiantes sp cifi es ATTENTION Tous les composants accessibles auront le m me potentiel que la fiche d entr e BNC ext rieure pour assurer la s curit...

Страница 10: ...nt les derni res 0 3s 1 mise jour 1s 2 mises jour par seconde ou 10s 1 mise jour par seconde selon le temps de mesure s lectionn Il faut noter que si un temps de mesure de 1s ou 10s est s lectionn e d...

Страница 11: ...ibilit de l entr e utilis e peut des mesures fausses l att nuation externe du signal avant de le mesurer pourra aider obtenir une mesure correcte Les signaux vitesses faibles tels que les signaux sinu...

Страница 12: ...au erhalb der angegebenen Nennversorgungsspannungen oder Umgebungsbedingungen betreiben WARNUNG Bei Betrieb des Ger ts mit den eingebauten Batterien oder einem freigegebenen Netzanschlussger t liegt a...

Страница 13: ...6 stellig angezeigt nach 1s 7 stellig und nach 10s 8 stellig falls die 10s Messung berhaupt gew hlt wurde Die Messfunktionen f r Frequenz und Periode erzeugen mindestens sieben Ziffern pro Sekunde Mes...

Страница 14: ...gnale mit langsam steigenden Flanken wie z B niederfrequente Sinuswellen fluktuieren aufgrund des wei en Rauschens etwas ZEITABLENKUNG UND ANDERE GENAUIGKEITSERW GUNGEN Das Ger t hat einen internen Os...

Страница 15: ...te o in condizioni ambientali non ammesse AVVISO Tutte le parti accessibili si trovano allo stesso potenziale delle parti esterne delle prese d ingresso BNC per garantire la sicurezza dell utente perc...

Страница 16: ...i ultimi 0 3s 1 aggiornamento 1s 2 aggiornamenti al secondo o 10s 1 aggiornamento al secondo a seconda del tempo di misurazione selezionato Nota se si seleziona un tempo di misurazione di 1s o 10s l a...

Страница 17: ...che si sta utilizzando possono provocare misurazioni sbagliate l attenuazione esterna del segnale prima che venga presentato al misuratore pu favorire una lettura corretta I segnali con velocit massi...

Страница 18: ...to no debe ser utilizado fuera de su clasificaci n de voltaje o de su gama ambiental ADVERTENCIA Todas las piezas accesibles estar n a la misma tensi n potencial que los enchufes BNC externos de entra...

Страница 19: ...ci n 1s 2 actualizaciones por segundo 10s 1 actualizaci n por segundo dependiendo del tiempo de medici n seleccionado Tenga en cuenta que si selecciona un tiempo de medici n de 1s 10s el inicio o rein...

Страница 20: ...idad de entrada usada pueden ayudar a obtener contajes incorrectos si se aten a la se al externamente antes de presentarla al contador se lograr una lectura correcta Las se ales con bajas velocidades...

Страница 21: ...idgeshire PE29 7DR England United Kingdom Telephone 44 0 1480 412451 Fax 44 0 1480 450409 International web site www aimtti com UK web site www aimtti co uk Email info aimtti com Aim Instruments and T...

Отзывы: