37
Chapter 5
Troubleshooting
TSI recommends that you return your S
IDE
P
AK
™
Model SP330 and SP350
Personal Sampling Pumps to the factory for annual service. Regular factory-
authorized cleaning and testing helps ensure that your instrument is working
properly, has the latest updates, and will provide reliable performance.
If you are having a problem with your S
IDE
P
AK
™
SP330 or SP350, use the
information below in the Troubleshooting matrix to try and resolve it in the
field. If necessary, contact TSI Incorporated or a local TSI distributor to
arrange for service.
Contact information:
TSI Incorporated
500 Cardigan Road
Shoreview, MN 55126
USA
Tel:
651-490-2811 or 1-800-874-2811
Website:
E-mail:
The following table lists the symptoms, possible causes, and recommended
solutions for common problems encountered with the S
IDE
P
AK
™
SP330 and
SP350 Personal Sampling Pumps.
Symptom
Possible Cause
Corrective Action
Nothing visible on
display.
Unit not switched on.
Low or dead batteries.
Batteries installed
backwards.
Dirty battery contacts.
Switch unit on.
Replace the batteries or
plug in the AC adapter.
Observe polarity
indication on battery
holder diagram.
Clean the battery
contacts.
Содержание SIDEPAK SP330
Страница 8: ......
Страница 31: ...Operation 23 See Appendix E for Quick Reference Guide...
Страница 44: ......
Страница 48: ......
Страница 55: ...47 Appendix B CSA Certificate of Compliance...
Страница 56: ...Appendix B 48...
Страница 57: ...CSA Certificate of Compliance 49...
Страница 58: ......
Страница 59: ...51 Appendix C LCIE ATEX Certificate of Compliance...
Страница 60: ...Appendix C 52...
Страница 61: ...CSA Certificate of Compliance 53...
Страница 62: ......
Страница 63: ...55 Appendix D EC Declaration of Conformity...
Страница 64: ......
Страница 65: ...57 Appendix E SP330 350 Quick Reference Guide See next page...
Страница 66: ...Appendix D 58...