Getting Started
2-3
Description
Function
Alarm
This connector provides two pins for a contact closure to control
an external alarm. The contact closure is normally open and
rated for 0 to 60 V AC/DC at 1.5A peak, 0.5A continuous. The
“alarm out” connection is rated for 60 V insulation. The contact
is closed under alarm conditions determined by the
programming of the device.
Filter scanner
This connector is used with the optional TSI Electronic Filter
Scanner (TSI P/N 700103).
Probe
This connector is used with a variety of TSI velocity /
temperature / humidity probes.
USB A
This standard USB connector is provided for use with “flash
drive” devices to download data from the instrument and
transfer to a computer or other device. The data files are in
ASCII format for easy use in spreadsheet programs.
This port can also be used to also connect a keyboard.
USB B
The standard USB connector is used to connect the instrument
to a computer running TSI T
RAK
P
RO
™ Lite software for data
downloading and recipe uploading. It is used with a standard
USB cable (included).
Ethernet port
The particle counter is compatible with either 10 or 100 MBps
systems. The green LED indicates that the network is connected.
The yellow LED indicates activity on the network cable.
Alarm
Ethernet
Port
Pump
Exhaust
Diffuser
Power
Connection
Battery
Access
Filter
Scanner
Probe
USB A
(Flash Drive)
USB B (to
computer)
Содержание 9510
Страница 2: ......
Страница 6: ......
Страница 18: ...1 6 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Страница 62: ...3 30 AEROTRAK Portable Airborne Particle Counter The standard reports are shown below ...
Страница 72: ...5 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...
Страница 76: ...6 4 AEROTRAK Portable Airborne Particle Counter This page intentionally left blank ...